Literature DB >> 22183896

Low pediatric cochlear implant failure rate: contributing factors in large-volume practice.

Antoine Eskander1, Karen A Gordon, Latif Kadhim, Vicky Papaioannou, Sharon L Cushing, Adrian L James, Blake C Papsin.   

Abstract

OBJECTIVE: To evaluate the rate of cochlear implant (CI) failure and CI reimplantation (CIri) in our population of children receiving implants by means of a technique that includes device fixation with suture to cortical bone.
DESIGN: Retrospective analysis from January 1990 to June 2010.
SETTING: Tertiary pediatric hospital. PATIENTS: A total of 971 devices were provided to 738 children (5575 implant-years).
INTERVENTIONS: Cochlear implant explants and CIri. MAIN OUTCOME MEASURES: Surgical findings at CIri were assessed by device model. The Pediatric Ranked Order Speech Perception score and the Phonetically Balanced Kindergarten score were used to make comparisons of hearing ability before and after CIri.
RESULTS: Thirty-four patients have undergone CIri in our pediatric center during the past 20 years. Excluding 7 of these patients who received their initial implant at other centers, our rate of CIri was 2.9%. Mean (SD) time to device failure was 61 (43) months. A disproportionately high number of patients (7 of 35 [20%]) requiring CIri had meningitis before implantation. After CIri, children maintained or improved their best speech performance measured before device failure, with only 2 children showing a significant reduction in speech perception after CIri.
CONCLUSIONS: A very low rate of failure occurs in children who receive CI devices, and several factors may account for this low rate. Children who develop meningitis before CI appear to be at an increased risk of device failure.

Entities:  

Mesh:

Year:  2011        PMID: 22183896     DOI: 10.1001/archoto.2011.200

Source DB:  PubMed          Journal:  Arch Otolaryngol Head Neck Surg        ISSN: 0886-4470


  8 in total

1.  Cochlear Implant Reliability: On the Reporting of Rates of Revision Surgery.

Authors:  Graham O'Neill; Neil S Tolley
Journal:  Indian J Otolaryngol Head Neck Surg       Date:  2020-04-01

Review 2.  Cochlear Implants and Children with Vestibular Impairments.

Authors:  Sharon L Cushing; Blake C Papsin
Journal:  Semin Hear       Date:  2018-07-20

3.  Cochlear re-implant rates in children: 20 years experience in a quaternary paediatric cochlear implant centre.

Authors:  Marilena Trozzi; Harry R F Powell; Shamim Toma; Waseem Ahmed; Christopher G Jephson; Kaukab Rajput; Lesley A Cochrane
Journal:  Eur Arch Otorhinolaryngol       Date:  2014-08-09       Impact factor: 2.503

4.  Clinical profile and management of revision cochlear implant surgeries.

Authors:  Ahmad M Aldhafeeri; Farid Alzhrani; Suliman Alajlan; Abdulrahman AlSanosi; Abdulrahman Hager
Journal:  Saudi Med J       Date:  2021-02       Impact factor: 1.484

5.  Implantable Devices for Single-Sided Deafness and Conductive or Mixed Hearing Loss: A Health Technology Assessment.

Authors: 
Journal:  Ont Health Technol Assess Ser       Date:  2020-03-06

Review 6.  Bilateral Cochlear Implantation: A Health Technology Assessment.

Authors: 
Journal:  Ont Health Technol Assess Ser       Date:  2018-10-24

Review 7.  Otolithic organ function in patients with profound sensorineural hearing loss.

Authors:  Yujuan Zhou; Yongzhen Wu; Jing Wang
Journal:  J Otol       Date:  2016-06-08

8.  Histopathological changes to the peripheral vestibular system following meningitic labyrinthitis.

Authors:  Henrique F Pauna; Renata M Knoll; Rory J Lubner; Jacob R Brodsky; Sharon L Cushing; Miguel A Hyppolito; Joseph B Nadol; Aaron K Remenschneider; Elliott D Kozin
Journal:  Laryngoscope Investig Otolaryngol       Date:  2020-02-14
  8 in total