An experimental study of electrochemical polishing for micro-electro-discharge-machined stainless-steel stents.

Derry Lappin1, Abdolreza Rashidi Mohammadi, Kenichi Takahata.   

Abstract

This paper reports electrochemical polishing (EP) of 316L stainless-steel structures patterned using micro-electro-discharge machining (μEDM) for application to stents including intelligent stents based on micro-electro-mechanical-systems technologies. For the process optimization, 10 μm deep cavities μEDMed on the planar material were polished in a phosphoric acid-based electrolyte with varying current densities and polishing times. The EP condition with a current density of 1.5 A/cm(2) for an EP time of 180 s exhibited the highest surface quality with an average roughness of 28 nm improved from~400 nm produced with high-energy μEDM. The EP of μEDMed surfaces was observed to produce almost constant smoothness regardless of the initial roughness determined by varying discharge energies. Energy-dispersive X-ray spectroscopy was performed on the μEDMed surfaces before and after EP. A custom rotational apparatus was used to polish tubular test samples including stent-like structures created using μEDM, demonstrating uniform removal of surface roughness and sharp edges from the structures.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 22183790     DOI: 10.1007/s10856-011-4513-2

Source DB:  PubMed          Journal:  J Mater Sci Mater Med        ISSN: 0957-4530            Impact factor:   3.896


  15 in total

Review 1.  Fluid mechanics of vascular systems, diseases, and thrombosis.

Authors:  D M Wootton; D N Ku
Journal:  Annu Rev Biomed Eng       Date:  1999       Impact factor: 9.590

2.  Electrochemical polishing of 316L stainless steel slotted tube coronary stents.

Authors:  Hui Zhao; Jan Van Humbeeck; Jürgen Sohier; Ivan De Scheerder
Journal:  J Mater Sci Mater Med       Date:  2002-10       Impact factor: 3.896

Review 3.  Review article: current management of primary sclerosing cholangitis.

Authors:  S N Cullen; R W Chapman
Journal:  Aliment Pharmacol Ther       Date:  2005-04-15       Impact factor: 8.171

Review 4.  Airway stenting.

Authors:  Yukihito Saito; Hiroji Imamura
Journal:  Surg Today       Date:  2005       Impact factor: 2.549

Review 5.  Biocompatibility aspects of new stent technology.

Authors:  O F Bertrand; R Sipehia; R Mongrain; J Rodés; J C Tardif; L Bilodeau; G Côté; M G Bourassa
Journal:  J Am Coll Cardiol       Date:  1998-09       Impact factor: 24.094

6.  Roughness and thrombogenicity of the outer surfaces of intravascular catheters.

Authors:  J F Hecker; L A Scandrett
Journal:  J Biomed Mater Res       Date:  1985-04

Review 7.  Use of stents in head and neck surgery.

Authors:  Ricardo L Carrau
Journal:  Curr Opin Otolaryngol Head Neck Surg       Date:  2005-04       Impact factor: 2.064

Review 8.  Sleep apnea and cardiovascular disease.

Authors:  Alexander G Logan; T Douglas Bradley
Journal:  Curr Hypertens Rep       Date:  2010-06       Impact factor: 5.369

Review 9.  Carotid artery stenting for the high surgical risk patients.

Authors:  N Al-Mubarak; S S Iyer
Journal:  J Cardiovasc Surg (Torino)       Date:  2005-02       Impact factor: 1.888

10.  Effect of surface roughness on platelet adhesion under static and under flow conditions.

Authors:  W Zingg; A W Neumann; A B Strong; O S Hum; D R Absolom
Journal:  Can J Surg       Date:  1982-01       Impact factor: 2.089

View more
  1 in total

1.  Characteristics of Phospholipid-Immunosuppressant-Antioxidant Mixed Langmuir-Blodgett Films.

Authors:  Małgorzata Jurak; Klaudia Szafran; Pilar Cea; Santiago Martín
Journal:  J Phys Chem B       Date:  2022-09-06       Impact factor: 3.466

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.