Automatic exposure control at MDCT based on the contrast-to-noise ratio: theoretical background and phantom study.

Yoshinori Funama1, Yoshiaki Sugaya, Osamu Miyazaki, Daisuke Utsunomiya, Yasuyuki Yamashita, Kazuo Awai.   

Abstract

PURPOSE: To develop a new automatic exposure control (AEC) technique based on the contrast-to-noise ratio (CNR) and provide constant lesion detectability.
METHODS: Lesion detectability is affected by factors such as image noise, lesion contrast, and lesion size. We performed ROC analysis to assess the relationship between the optimum CNR and the lesion diameter at various levels of lesion contrast. We then developed a CNR-based AEC algorithm based on lesion detectability. Using CNR- based AEC algorithm, we performed visual evaluation of low-contrast detectability by 5 radiologists on a low-contrast module of the Catphan phantom, a contrast-difference level of 1.0% (difference in the CT number = 10 HU), and objects 3.0-9.0 mm in diameter.
RESULTS: On step-and-shoot scans the mean detection fraction with CNR-based AEC remained almost constant from 88 to 99 % regardless of the lesion size. We observed the same trend on helical scans, the mean detection fraction with CNR-based AEC exhibited a high score from 91 to 100%. Although CNR-based AEC maintains higher CNR for smaller size or lower contrast lesion, radiation dose on 3 mm lesion resulted in about 13 times larger than that of 9 mm lesion size. CTDI(vol) for the CNR-based AEC technique changed dramatically with the SD(Z) from 7.5 to 100.0 mGy for step-and-shoot scans and from 9.1 to 121.5 mGy for helical scans.
CONCLUSIONS: From the viewpoint of ROC analysis-based CNR for lesion detection, CNR-based AEC potentially provide image quality advantages for clinical implementation.
Copyright © 2011 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

Mesh:

Year:  2011        PMID: 22182517     DOI: 10.1016/j.ejmp.2011.11.004

Source DB:  PubMed          Journal:  Phys Med        ISSN: 1120-1797            Impact factor:   2.685


  4 in total

1.  Optimization of dental CBCT exposures through mAs reduction.

Authors:  R Pauwels; L Seynaeve; J C G Henriques; C de Oliveira-Santos; P C Souza; F H Westphalen; I R F Rubira-Bullen; R F Ribeiro-Rotta; M I B Rockenbach; F Haiter-Neto; P Pittayapat; H Bosmans; R Bogaerts; R Jacobs
Journal:  Dentomaxillofac Radiol       Date:  2015-06-19       Impact factor: 2.419

2.  Radiation dose reduction based on CNR index with low-tube voltage scan for pediatric CT scan: experimental study using anthropomorphic phantoms.

Authors:  Takanori Masuda; Yoshinori Funama; Masao Kiguchi; Naoyuki Imada; Takayuki Oku; Tomoyasu Sato; Kazuo Awai
Journal:  Springerplus       Date:  2016-12-01

3.  CT image quality over time: comparison of image quality for six different CT scanners over a six-year period.

Authors:  Ana Maria A Roa; Hilde K Andersen; Anne Catrine T Martinsen
Journal:  J Appl Clin Med Phys       Date:  2015-03-08       Impact factor: 2.102

4.  Relationship between size-specific dose estimates and image quality in computed tomography depending on patient size.

Authors:  Hiroki Kawashima; Katsuhiro Ichikawa; Shinsuke Hanaoka; Kosuke Matsubara; Tadanori Takata
Journal:  J Appl Clin Med Phys       Date:  2018-05-04       Impact factor: 2.102

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.