Ultrasound assessment of inferior vena cava collapsibility is not a valid measure of preload changes during triggered positive pressure ventilation: a controlled cross-over study.

P Juhl-Olsen1, C A Frederiksen, E Sloth.   

Abstract

PURPOSE: Respiratory changes in the diameter of the inferior vena cava (IVC) have been validated as a measure of volume status and preload responsiveness during spontaneous breathing and mechanical ventilation. However, many intensive care patients are ventilated with triggered positive pressure ventilation (PPV). In this setting, there is no evidence regarding IVC collapsibility (IVCc) as a surrogate for preload. We aimed to elucidate the effects of increasing levels of triggered PPV and of varying preload conditions on the IVCc.
MATERIALS AND METHODS: 10 healthy volunteers were connected to a ventilator through a tight-fitting mask and exposed to 6 different levels of positive end-expiratory pressure (PEEP) and pressure support (PS) after a baseline reading. All ventilator settings were performed at neutral preload (horizontal position), low preload (reverse-Trendelenburg) and high preload (Trendelenburg position with an intravenous fluid bolus). At each ventilator setting, the IVC was imaged throughout at least 1 respiratory cycle using 3 commonly used ultrasound techniques including sagittal M-mode and 2-dimensional echocardiography in both sagittal and transverse views.
RESULTS: Increasing PS diminished IVCc (p = 0.01) in the reverse-Trendelenburg position, and increasing PEEP caused a higher IVCc in the Trendelenburg position (p = 0.02). In the horizontal position, no significant effects of increasing PS, PEEP or a combination of the two were seen. Overall ANOVA analysis showed that IVCc was not independent of preload. During PPV, IVCc was highest at neutral preload at most ventilator settings, IVCc was lowest at low preload, while high preload generally facilitated an IVCc between neutral and high preload. In addition, sagittal M-mode and transverse 2-dimensional echocardiography overestimated IVCc as compared to sagittal 2-dimensional echocardiography.
CONCLUSION: The compiled results of this study show that IVCc cannot be held as a valid measure of preload status during PPV. This may be explained by systematic alterations in other determinants for IVCc. Comparison of methods encourages the use of sagittal 2-dimensional echocardiography for dynamic imaging of the IVC. Sagittal M-mode and transverse 2-dimensional echocardiography overestimate IVCc as compared to sagittal 2-dimensional echocardiography. © Georg Thieme Verlag KG Stuttgart · New York.

Entities:  

Mesh:

Year:  2011        PMID: 22179799     DOI: 10.1055/s-0031-1281832

Source DB:  PubMed          Journal:  Ultraschall Med        ISSN: 0172-4614            Impact factor:   6.548


  6 in total

Review 1.  Cardiopulmonary interactions and volume status assessment.

Authors:  Alain F Broccard
Journal:  J Clin Monit Comput       Date:  2012-08-30       Impact factor: 2.502

2.  Ten situations where inferior vena cava ultrasound may fail to accurately predict fluid responsiveness: a physiologically based point of view.

Authors:  G Via; G Tavazzi; S Price
Journal:  Intensive Care Med       Date:  2016-04-23       Impact factor: 17.440

Review 3.  [Hemodynamic monitoring in intensive care and emergency medicine : Integration of clinical signs and ultrasound findings].

Authors:  D Hempel; R Pfister; G Michels
Journal:  Med Klin Intensivmed Notfmed       Date:  2016-06-09       Impact factor: 0.840

Review 4.  Point-of-Care Ultrasound for the Trauma Anesthesiologist.

Authors:  Eric R Heinz; Anita Vincent
Journal:  Curr Anesthesiol Rep       Date:  2022-01-20

5.  The relationship between inferior vena cava diameter measured by bedside ultrasonography and central venous pressure value.

Authors:  Serenat Citilcioglu; Ahmet Sebe; Mehmet Oguzhan Ay; Ferhat Icme; Akkan Avci; Muge Gulen; Mustafa Sahan; Salim Satar
Journal:  Pak J Med Sci       Date:  2014-03       Impact factor: 1.088

6.  Measurement of the vascular pedicle width predicts fluid repletion: a cross-sectional comparison with inferior vena cava ultrasound and lung comets.

Authors:  Nawal Salahuddin; Iqbal Hussain; Hakam Alsaidi; Quratulain Shaikh; Mini Joseph; Hassan Hawa; Khalid Maghrabi
Journal:  J Intensive Care       Date:  2015-12-22
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.