Bubble CPAP versus CPAP with variable flow in newborns with respiratory distress: a randomized controlled trial.

Ana Cristina Zanon Yagui1, Luciana Assis Pires Andrade Vale, Luciana Branco Haddad, Cristiane Prado, Felipe Souza Rossi, Alice D Agostini Deutsch, Celso Moura Rebello.   

Abstract

OBJECTIVE: To evaluate the efficacy and safety of nasal continuous positive airway pressure (NCPAP) using devices with variable flow or bubble continuous positive airway pressure (CPAP) regarding CPAP failure, presence of air leaks, total CPAP and oxygen time, and length of intensive care unit and hospital stay in neonates with moderate respiratory distress (RD) and birth weight (BW) ≥ 1,500 g.
METHODS: Forty newborns requiring NCPAP were randomized into two study groups: variable flow group (VF) and continuous flow group (CF). The study was conducted between October 2008 and April 2010. Demographic data, CPAP failure, presence of air leaks, and total CPAP and oxygen time were recorded. Categorical outcomes were tested using the chi-square test or the Fisher's exact test. Continuous variables were analyzed using the Mann-Whitney test. The level of significance was set at p < 0.05.
RESULTS: There were no differences between the groups with regard to demographic data, CPAP failure (21.1 and 20.0% for VF and CF, respectively; p = 1.000), air leak syndrome (10.5 and 5.0%, respectively; p = 0.605), total CPAP time (median: 22.0 h, interquartile range [IQR]: 8.00-31.00 h and median: 22.0 h, IQR: 6.00-32.00 h, respectively; p = 0.822), and total oxygen time (median: 24.00 h, IQR: 7.00-85.00 h and median: 21.00 h, IQR: 9.50-66.75 h, respectively; p = 0.779).
CONCLUSION: In newborns with BW ≥ 1,500 g and moderate RD, the use of continuous flow NCPAP showed the same benefits as the use of variable flow NCPAP.

Entities:  

Mesh:

Year:  2011        PMID: 22170173     DOI: 10.2223/JPED.2145

Source DB:  PubMed          Journal:  J Pediatr (Rio J)        ISSN: 0021-7557            Impact factor:   2.197


  6 in total

1.  Impact of the systematic introduction of low-cost bubble nasal CPAP in a NICU of a developing country: a prospective pre- and post-intervention study.

Authors:  Rossano Rezzonico; Letizia M Caccamo; Valeria Manfredini; Massimo Cartabia; Nieves Sanchez; Zoraida Paredes; Patrizia Froesch; Franco Cavalli; Maurizio Bonati
Journal:  BMC Pediatr       Date:  2015-03-25       Impact factor: 2.125

2.  Decreasing Chronic Lung Disease Associated with Bubble CPAP Technology: Experience at Five Years.

Authors:  Tricia A Miller; Jing Li; Stella Riddell; Steven C Barkley
Journal:  Pediatr Qual Saf       Date:  2020-04-10

3.  Bubble CPAP therapy for neonatal respiratory distress in level III neonatal unit in Amman, Jordan: a prospective observational study.

Authors:  Manar Al-Lawama; Haitham Alkhatib; Zaid Wakileh; Randa Elqaisi; Ghada AlMassad; Eman Badran; Tyler Hartman
Journal:  Int J Gen Med       Date:  2018-12-24

Review 4.  Clinical review of High Flow Nasal Cannula and Continuous Positive Airway Pressure in pediatric acute respiratory distress.

Authors:  Kurniawan Taufiq Kadafi; Saptadi Yuliarto; Charity Monica; William Prayogo Susanto
Journal:  Ann Med Surg (Lond)       Date:  2021-12-14

5.  Effectiveness of Bubble Continuous Positive Airway Pressure (BCPAP) for Treatment of Children Aged 1-59 Months with Severe Pneumonia and Hypoxemia in Ethiopia: A Pragmatic Cluster Randomized Controlled Clinical Trial.

Authors:  Meseret Gebre; Kassa Haile; Trevor Duke; Md Tanveer Faruk; Mehnaz Kamal; Md Farhad Kabir; Md Fakhar Uddin; Muluye Shimelis; Bethelhem Solomon; Abebe Genetu Bayih; Alemseged Abdissa; Taye Tolera Balcha; Rahel Argaw; Asrat Demtse; Abate Yeshidenber; Abayneh Girma; Bitseat W Haile; Tahmeed Ahmed; John D Clemens; Mohammod Jobayer Chisti
Journal:  J Clin Med       Date:  2022-08-23       Impact factor: 4.964

Review 6.  Practical aspects on the use of non-invasive respiratory support in preterm infants.

Authors:  Nehad Nasef; Hend Me Rashed; Hany Aly
Journal:  Int J Pediatr Adolesc Med       Date:  2020-02-18