| Literature DB >> 22163440 |
Hirotaka Sakaue1, Keiko Ishii.
Abstract
Anodized-aluminum pressure-sensitive paint (AA-PSP) uses the dipping deposition method to apply a luminophore on a porous anodized-aluminum surface. We study the dipping duration, one of the parameters of the dipping deposition related to the characterization of AA-PSP. The dipping duration was varied from 1 to 100,000 s. The properties characterized are the pressure sensitivity, temperature dependency, and signal level. The maximum pressure sensitivity of 65% is obtained at the dipping duration of 100 s, the minimum temperature dependency is obtained at the duration of 1 s, and the maximum signal level is obtained at the duration of 1,000 s, respectively. Among the characteristics, the dipping duration most influences the signal level. The change in the signal level is a factor of 8.4. By introducing a weight coefficient, an optimum dipping duration can be determined. Among all the dipping parameters, such as the dipping duration, dipping solvent, and luminophore concentration, the pressure sensitivity and signal level are most influenced by the dipping solvent.Entities:
Keywords: anodized aluminum; dipping duration; pressure-sensitive paint
Mesh:
Substances:
Year: 2010 PMID: 22163440 PMCID: PMC3230996 DOI: 10.3390/s101109799
Source DB: PubMed Journal: Sensors (Basel) ISSN: 1424-8220 Impact factor: 3.576
Dipping conditions of AA-PSP. Bathophen ruthenium was chosen as a luminophore and dichloromethane was chosen as a solvent. The luminophore concentration was fixed at 0.1 mM.
| AAPSP1 | 1 |
| AAPSP10 | 10 |
| AAPSP100 | 100 |
| AAPSP1000 | 1,000 |
| AAPSP3600 | 3,600 |
| AAPSP100000 | 100,000 |
Figure 1.Pressure calibrations of AA-PSPs with varying the dipping duration.
Summary of AA-PSP characterizations. The error was determined as the standard deviation of the three data sets from the same dipping procedures.
| AAPSP1 | 52 ± 1 | −1.10 ± 0.10 | 20.0 ± 2.8 |
| AAPSP10 | 56 ± 2 | −1.63 ± 0.30 | 45.2 ± 6.8 |
| AAPSP100 | 65 ± 2 | −2.35 ± 0.22 | 139.4 ± 20.4 |
| AAPSP1000 | 63 ± 2 | −2.11 ± 0.11 | 167.1 ± 17.1 |
| AAPSP3600 | 62 ± 4 | −1.38 ± 0.15 | 100.0 ± 14.8 |
| AAPSP100000 | 59 ± 4 | −1.28 ± 0.21 | 67.2 ± 11.2 |
Figure 2.Temperature calibrations of AA-PSPs with varying the dipping duration.
Figure 3.Normalized outputs of AA-PSP. The pressure sensitivity is shown as normσ, the temperature dependency as normδ, and the signal level as normη, respectively.
Optimum value, n, determined from weight coefficients, α, α, and α for given dipping duration. Condition *1: α = 1 and others are zero. Condition *2: α = α = α = 1/3.
| AAPSP1 | 0.00 | 0.33 |
| AAPSP10 | 0.30 | 0.34 |
| AAPSP100 | 1.00 | 0.60 |
| AAPSP1000 | 0.85 | 0.68 |
| AAPSP3600 | 0.82 | 0.71 |
| AAPSP100000 | 0.56 | 0.58 |
The maximum and minimum AA-PSP characterizations reported from Refs. [2,3]. The results from the present case were also listed.
| 62 | 6 | 62 | 31 | 65 | 52 | |
| NA | NA | −0.62 | −1.44 | −1.10 | −2.35 | |
| 189 | 13 | 100 | 27.5 | 167.1 | 20.1 | |