Abstract

BACKGROUND: Proper tip position is a major determinant of totally implantable venous access device (TIVAD) outcome. The aim of this study is to analyze the potential utilization of intravenous electrocardiography (IV-ECG) to help inexperienced operators for TIVAD placement. PATIENTS AND METHODS: This is a retrospective, observational, uni-institutional study. 331 patients receiving TIVAD implantation from July 2008 to December 2008 were recruited. In IV-ECG group, IV-ECG was used to help decide catheter tip location and catheter length. In Landmark group, catheter length was decided by surface landmarks. Catheter tip position was confirmed by post-operative supine chest X-ray.
RESULTS: There were 153 patients in IV-ECG group, and 178 patients in Landmark group. No immediate reoperation due to catheter mal-position was noted in IV-ECG group, but it happened in eight patients in Landmark group. In IV-ECG group, 97.3% of the catheter tip located at proper position (within 2 cm from junction of right atrium and superior vena cava, as compared to 88.8% of the tip position in Landmark group was proper. The difference was statistically significant (P < 0.05). There was no complication associated with the use of IV-ECG.
CONCLUSION: IV-ECG is a safe and convenient method to help inexperienced operators placing TIVAD.
Copyright © 2011 Wiley Periodicals, Inc.

Entities:  

Mesh:

Year:  2011        PMID: 22161950     DOI: 10.1002/jso.23000

Source DB:  PubMed          Journal:  J Surg Oncol        ISSN: 0022-4790            Impact factor:   3.454


  2 in total

1.  Long-term outcomes of totally implantable venous access devices.

Authors:  Yi-Chia Wang; Pei-Lin Lin; Wei-Han Chou; Chih-Peng Lin; Chi-Hsiang Huang
Journal:  Support Care Cancer       Date:  2017-02-09       Impact factor: 3.603

2.  Transesophageal echocardiography-guided implantation of totally implantable venous access devices via the internal jugular vein: retrospective analysis of 297 cases in pediatric patients.

Authors:  Yuanzhen Chen; Dajun Xing; Lixin Wu; Huatian Lin; Ting Lin; Fang Ding; Liang Xu
Journal:  World J Surg Oncol       Date:  2022-08-30       Impact factor: 3.253

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.