Radiation leakage through electron applicators on Clinac-1800 accelerators.

A Perec1, H Kubo.   

Abstract

Radiation leakage through electron applicators by 6-, 9-, 12-, 16-, and 20-MeV electron beams from Varian Clinac-1800 has been measured with films. High levels of leakage were found under the corners of applicators and the touch-plate mounting port holes. The radiation leakage, relative to the central-axis dose at dmax, was found to be up to 13% on the patient plane [100-cm source-to-film distance (SFD)] and up to 35% beneath the corners of applicators (96-cm SFD). Up to 18% radiation leakage was measured beneath the touch plate near the mounting port holes (96-cm SFD). The extent of radiation leakage in all electron beams was investigated and some simple shielding solutions to reduce the leakage are suggested.

Entities:  

Mesh:

Year:  1990        PMID: 2215419     DOI: 10.1118/1.596472

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  6 in total

1.  Quantification and reduction of peripheral dose from leakage radiation on Siemens Primus accelerators in electron therapy mode.

Authors:  Collins Yeboah; Alex Karotki; Dylan Hunt; Rick Holly
Journal:  J Appl Clin Med Phys       Date:  2010-06-15       Impact factor: 2.102

2.  Experimental assessment of out-of-field dose components in high energy electron beams used in external beam radiotherapy.

Authors:  Mohamad M Alabdoaburas; Jean-Pierre Mege; Jean Chavaudra; Jérémi Vũ Bezin; Atilla Veres; Florent de Vathaire; Dimitri Lefkopoulos; Ibrahima Diallo
Journal:  J Appl Clin Med Phys       Date:  2015-11-08       Impact factor: 2.102

3.  Verification of electron beam parameters in an intraoperative linear accelerator using dosimetric and radiobiological response methods.

Authors:  Agnieszka Dróżdż; Martyna Waluś; Marcin Zieliński; Bożena Malesa; Marta Kruszyna-Mochalska; Katarzyna Kulcenty; Beata Adamczyk; Piotr Nowaczyk; Julian Malicki; Jacek Pracz
Journal:  Rep Pract Oncol Radiother       Date:  2021-12-30

4.  Nontarget and Out-of-Field Doses from Electron Beam Radiotherapy.

Authors:  Natalia Matuszak; Marta Kruszyna-Mochalska; Agnieszka Skrobala; Adam Ryczkowski; Piotr Romanski; Igor Piotrowski; Katarzyna Kulcenty; Wiktoria Maria Suchorska; Julian Malicki
Journal:  Life (Basel)       Date:  2022-06-08

5.  Out-of-field doses and neutron dose equivalents for electron beams from modern Varian and Elekta linear accelerators.

Authors:  Carlos E Cardenas; Paige L Nitsch; Rajat J Kudchadker; Rebecca M Howell; Stephen F Kry
Journal:  J Appl Clin Med Phys       Date:  2016-07-08       Impact factor: 2.102

6.  Radiation leakage dose from Elekta electron collimation system.

Authors:  Garrett M Pitcher; Kenneth R Hogstrom; Robert L Carver
Journal:  J Appl Clin Med Phys       Date:  2016-09-08       Impact factor: 2.102

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.