Björn Gamm1, Holger Blank, Radian Popescu, Reinhard Schneider, André Beyer, Armin Gölzhäuser, Dagmar Gerthsen.
Abstract
Single atoms can be considered as the most basic objects for electron microscopy to test the microscope performance and basic concepts for modeling image contrast. In this work high-resolution transmission electron microscopy was applied to image single platinum, molybdenum, and titanium atoms in an aberration-corrected transmission electron microscope. The atoms are deposited on a self-assembled monolayer substrate that induces only negligible contrast. Single-atom contrast simulations were performed on the basis of Weickenmeier-Kohl and Doyle-Turner form factors. Experimental and simulated image intensities are in quantitative agreement on an absolute intensity scale, which is provided by the vacuum image intensity. This demonstrates that direct testing of basic properties such as form factors becomes feasible.Entities:
Year: 2011 PMID: 22153521 DOI: 10.1017/S1431927611012232
Source DB: PubMed Journal: Microsc Microanal ISSN: 1431-9276 Impact factor: 4.127