Thirty-minute compared to standardised office blood pressure measurement in general practice.

Nynke Scherpbier-de Haan1, Mark van der Wel, Gijs Schoenmakers, Steve Boudewijns, Petronella Peer, Chris van Weel, Theo Thien, Carel Bakx.   

Abstract

BACKGROUND: Although blood pressure measurement is one of the most frequently performed measurements in clinical practice, there are concerns about its reliability. Serial, automated oscillometric blood pressure measurement has the potential to reduce measurement bias and white-coat effect'. AIM: To study agreement of 30-minute office blood pressure measurement (OBPM) with standardised OBPM, and to compare repeatability. DESIGN AND
SETTING: Method comparison study in two general practices in The Netherlands.
METHOD: Thirty-minute and standardised OBPM was carried out with the same, validated device in 83 adult patients, and the procedure was repeated after 2 weeks. During 30-minute OBPM, blood pressure was measured automatically every 3 minutes, with the patient in a sitting position, alone in a quiet room. Agreement between 30-minute and standardised OBPM was assessed by Bland-Altman analysis. Repeatability of the blood pressure measurement methods after 2 weeks was expressed as the mean difference in combination with the standard deviation of difference (SDD).
RESULTS: Mean 30-minute OBPM readings were 7.6/2.5 mmHg (95% confidence interval [CI] = 6.1 to 9.1/1.5 to 3.4 mmHg) lower than standardised OBPM readings. The mean difference and SDD between repeated 30-minute OBPMs (mean difference = 3/1 mmHg, 95% CI = 1 to 5/0 to 2 mmHg; SDD 9.5/5.3 mmHg) were lower than those of standardised OBPMs (mean difference = 6/2 mmHg, 95% CI = 4 to 8/1 to 4 mmHg; SDD 10.9/6.3 mmHg).
CONCLUSION: Thirty-minute OBPM resulted in lower readings than standardised OBPM and had a better repeatability. These results suggest that 30-minute OBPM better reflects the patient's true blood pressure than standardised OBPM does.

Entities:  

Mesh:

Year:  2011        PMID: 22152748      PMCID: PMC3162182          DOI: 10.3399/bjgp11X593875

Source DB:  PubMed          Journal:  Br J Gen Pract        ISSN: 0960-1643            Impact factor:   5.386


  26 in total

Review 1.  Use and interpretation of ambulatory blood pressure monitoring: recommendations of the British hypertension society.

Authors:  E O'Brien; A Coats; P Owens; J Petrie; P L Padfield; W A Littler; M de Swiet; F Mee
Journal:  BMJ       Date:  2000-04-22

2.  A validation of the Mobil O Graph (version 12) ambulatory blood pressure monitor.

Authors:  C R Jones; K Taylor; P Chowienczyk; L Poston; A H Shennan
Journal:  Blood Press Monit       Date:  2000-08       Impact factor: 1.444

3.  How should we measure blood pressure in the doctor's office?

Authors:  W Gerin; R M Marion; R Friedman; G D James; D H Bovbjerg; T G Pickering
Journal:  Blood Press Monit       Date:  2001-10       Impact factor: 1.444

4.  Comparison of acceptability of and preferences for different methods of measuring blood pressure in primary care.

Authors:  Paul Little; Jane Barnett; Lucy Barnsley; Jean Marjoram; Alex Fitzgerald-Barron; David Mant
Journal:  BMJ       Date:  2002-08-03

5.  Reliability of self-reported blood pressure measurements.

Authors:  K A Johnson; D J Partsch; L L Rippole; D M McVey
Journal:  Arch Intern Med       Date:  1999 Dec 13-27

6.  Reproducibility of home, ambulatory, and clinic blood pressure: implications for the design of trials for the assessment of antihypertensive drug efficacy.

Authors:  George S Stergiou; Nikolaos M Baibas; Alexandra P Gantzarou; Irini I Skeva; Chrysa B Kalkana; Leonidas G Roussias; Theodore D Mountokalakis
Journal:  Am J Hypertens       Date:  2002-02       Impact factor: 2.689

Review 7.  Blood Pressure Monitoring. Task force V: White-coat hypertension.

Authors:  T G Pickering; A Coats; J M Mallion; G Mancia; P Verdecchia
Journal:  Blood Press Monit       Date:  1999-12       Impact factor: 1.444

8.  Recommendations for blood pressure measurement in humans and experimental animals: part 1: blood pressure measurement in humans: a statement for professionals from the Subcommittee of Professional and Public Education of the American Heart Association Council on High Blood Pressure Research.

Authors:  Thomas G Pickering; John E Hall; Lawrence J Appel; Bonita E Falkner; John Graves; Martha N Hill; Daniel W Jones; Theodore Kurtz; Sheldon G Sheps; Edward J Roccella
Journal:  Circulation       Date:  2005-02-08       Impact factor: 29.690

9.  Comparison of agreement between different measures of blood pressure in primary care and daytime ambulatory blood pressure.

Authors:  Paul Little; Jane Barnett; Lucy Barnsley; Jean Marjoram; Alex Fitzgerald-Barron; David Mant
Journal:  BMJ       Date:  2002-08-03

10.  Conventional versus automated measurement of blood pressure in primary care patients with systolic hypertension: randomised parallel design controlled trial.

Authors:  Martin G Myers; Marshall Godwin; Martin Dawes; Alexander Kiss; Sheldon W Tobe; F Curry Grant; Janusz Kaczorowski
Journal:  BMJ       Date:  2011-02-07
View more
  9 in total

1.  Measuring blood pressure in primary care: identifying 'white coat syndrome' and blood pressure device comparison.

Authors:  Emma Wallace; Tom Fahey
Journal:  Br J Gen Pract       Date:  2011-09       Impact factor: 5.386

2.  Thirty-minute compared to standardised office blood pressure measurement in general practice.

Authors:  Akke Vellinga; Andrew Murphy; Eoin O'Brien
Journal:  Br J Gen Pract       Date:  2012-03       Impact factor: 5.386

3.  It is Time to Change How We Measure Blood Pressures in the Office.

Authors:  Lee A Green
Journal:  Ann Fam Med       Date:  2017-03       Impact factor: 5.166

4.  Implementation of an At-home Blood Pressure Measurement Protocol in a Hypertension Management Clinical Trial During the COVID-19 Pandemic.

Authors:  Hailey N Miller; Miriam B Berger; Sandy Askew; Elizabeth Trefney; Crystal Tyson; Laura Svetkey; Gary G Bennett; Dori M Steinberg
Journal:  J Cardiovasc Nurs       Date:  2022-04-23       Impact factor: 2.468

5.  Thirty-Minute Office Blood Pressure Monitoring in Primary Care.

Authors:  Michiel J Bos; Sylvia Buis
Journal:  Ann Fam Med       Date:  2017-03       Impact factor: 5.166

Review 6.  Sources of inaccuracy in the measurement of adult patients' resting blood pressure in clinical settings: a systematic review.

Authors:  Noa Kallioinen; Andrew Hill; Mark S Horswill; Helen E Ward; Marcus O Watson
Journal:  J Hypertens       Date:  2017-03       Impact factor: 4.844

7.  Comparison of multiple blood pressure frequency methods with optimum blood pressure measurement among Iranian individuals.

Authors:  Alireza Mohammadi-Dolatabadi; Mehrbod Vakhshoori; Maryam Eghbali-Babadi; Maryam Heidarpour; Davood Shafie; Mohammad Garakyaraghi; Alireza Khosravi
Journal:  J Res Med Sci       Date:  2020-04-13       Impact factor: 1.852

8.  Development and Psychometric Validation of the Hypertension Beliefs Assessment Tool Among Adult Population in Northwest Ethiopia.

Authors:  Destaw Fetene Teshome; Shitaye Alemu Balcha; Tadesse Awoke Ayele; Asmamaw Atnafu; Kassahun Alemu Gelaye
Journal:  Patient Prefer Adherence       Date:  2021-11-25       Impact factor: 2.711

9.  Comparison of the average of five readings with averages from fewer readings for automated oscillometric blood pressure measurement in an outpatient clinic.

Authors:  Idris Oladipo; Ayoade Adedokun
Journal:  Korean Circ J       Date:  2013-05-31       Impact factor: 3.243

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.