Case identification of subjects with airflow limitations using the handheld spirometer "Hi-Checker™" : comparison against an electronic desktop spirometer.

Koichi Nishimura1, Kazuhito Nakayasu, Atsuko Kobayashi, Satoshi Mitsuma.   

Abstract

BACKGROUND: Systematic case identification has been proposed as a strategy to improve diagnosis rates and to enable the early detection of subjects with COPD. We hypothesized that case identification could be possible using the handheld spirometer Hi-Checker™. AIM: To determine how to modify the FEV(1)/FEV(6) values obtained using the Hi-Checker™ to screen for cases with airflow limitation.
METHODS: Spirometry was performed with both an electronic desktop spirometer and with the Hi-Checker™ in 312 male subjects.
RESULTS: The average FEV(1) (mean ± SD) measured using a conventional spirometer and the Hi-Checker™ was 2.99 ± 0.56L and 3.07 ± 0.57L, respectively. These results were significantly different (P<0.001, paired t-test for both). This difference of -0.08 ± 0.13L (95% confidence interval: -0.094-0.066L) was normally distributed, and thought to be random. Statistically significant correlations were found for all measurements between the spirometer and the Hi-Checker™ ; the Pearson's correlation coefficient (R) between the FEV(1)/FVC and FEV(1)/FEV(6) values was 0.881. If one defines a FEV(1/)FVC smaller than 0.7 measured by the spirometer as airflow limitation, then a FEV(1)/FEV(6) smaller than 0.746 measured by the Hi-Checker™ best matches this definition, and Cohen's kappa coefficient was 0.672.
CONCLUSION: Although the Hi-Checker™ estimates resembled those from conventional spirometry, it should be emphasized that the two methods did not produce identical results due to random measurement error. Although one must be careful about overinterpreting these results, since the Hi-Checker™ is small and inexpensive, it could make a significant contribution in facilitating the case selection of patients with airflow limitation.

Entities:  

Mesh:

Year:  2011        PMID: 22149406     DOI: 10.3109/15412555.2011.626817

Source DB:  PubMed          Journal:  COPD        ISSN: 1541-2563            Impact factor:   2.409


  8 in total

Review 1.  Accuracy of portable spirometers in the diagnosis of chronic obstructive pulmonary disease A meta-analysis.

Authors:  Na Yu; Wei Wang; Jiawei Zhou; Xiaomeng Li; Xingjian Wang
Journal:  NPJ Prim Care Respir Med       Date:  2022-04-19       Impact factor: 3.289

2.  VitalQPlus: a potential screening tool for early diagnosis of COPD.

Authors:  Chee Fai Sui; Long Chiau Ming; Chin Fen Neoh; Baharudin Ibrahim
Journal:  Int J Chron Obstruct Pulmon Dis       Date:  2015-08-11

3.  The relationship between the COPD Assessment Test score and airflow limitation in Japan in patients aged over 40 years with a smoking history.

Authors:  Daisuke Yoshimoto; Yasutaka Nakano; Katsuya Onishi; Gerry Hagan; Paul Jones
Journal:  Int J Chron Obstruct Pulmon Dis       Date:  2014-12-09

4.  Early Detection of Chronic Obstructive Pulmonary Disease in Primary Care.

Authors:  Seiichi Kobayashi; Masakazu Hanagama; Masaru Yanai
Journal:  Intern Med       Date:  2017-09-25       Impact factor: 1.271

5.  The accuracy of a handheld "disposable pneumotachograph device" in the spirometric diagnosis of airway obstruction in a Chinese population.

Authors:  Guojun Chen; Longyuan Jiang; Liwen Wang; Wei Zhang; Carlos Castillo; Xiangshao Fang
Journal:  Int J Chron Obstruct Pulmon Dis       Date:  2018-08-02

6.  Validity of a portable spirometer in the communities of China.

Authors:  Shan Xiao; Fan Wu; Zihui Wang; Jianmin Chen; Huajing Yang; Youlan Zheng; Zhishan Deng; Jieqi Peng; Xiang Wen; Peiyu Huang; Cuiqiong Dai; Lifei Lu; Ningning Zhao; Pixin Ran; Yumin Zhou
Journal:  BMC Pulm Med       Date:  2022-03-05       Impact factor: 3.317

7.  Validity of the Handheld Expiratory Flowmeter for COPD Screening in the Primary Care Setting of China.

Authors:  Shuyun Chen; Xiaochen Li; Zihui Wang; Yumin Zhou; Dongxing Zhao; Zhuxiang Zhao; Sha Liu; Pixin Ran
Journal:  Int J Chron Obstruct Pulmon Dis       Date:  2021-07-08

8.  Detection of airflow limitation using a handheld spirometer in a primary care setting.

Authors:  Siew-Mooi Ching; Yong-Kek Pang; David Price; Ai-Theng Cheong; Ping-Yein Lee; Ismail Irmi; Hassan Faezah; Ismail Ruhaini; Yook-Chin Chia
Journal:  Respirology       Date:  2014-04-07       Impact factor: 6.424

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.