Guidewire stiffness: what's in a name?

Gareth J Harrison1, Thien V How, S Rao Vallabhaneni, John A Brennan, Robert K Fisher, Jagjeeth B Naik, Richard G McWilliams.   

Abstract

PURPOSE: To measure the stiffness of commonly used "stiff" guidewires in terms of their flexural modulus, an engineering parameter related to bending stiffness.
METHODS: Eleven different intact stiff guidewires were selected to undergo a 3-point bending test performed using a tensile testing machine. Testing was performed on 3 new and intact specimens of each guidewire at 10 locations along the wire's length, excluding the floppy tip. The flexural modulus (in gigapascals, GPa) was calculated from the results of the bending test.
RESULTS: The flexural modulus of the plain Amplatz wire was 9.5 GPa compared to 11.4 to 14.5 GPa for the "heavy duty" wires. Within the Amplatz family of guidewires, the flexural modulus was 17 GPa for the "stiff," 29.2 GPa for the "extra stiff," 60.3 GPa for the "super stiff," and 65.4 GPa for the "ultra stiff." The Backup Meier measured 139.6 GPa and the Lunderquist Extra Stiff 158.4 GPa.
CONCLUSION: The Instructions for Use of some endovascular devices specify a wire type selected from a range of undefined "stiffness" descriptors. These descriptors have little correlation with the measured flexural modulus. Two guidewires with the description "extra stiff" can have a 5-fold difference in flexural modulus. We recommend that guidewire catalogues and packaging include the flexural modulus and that device manufacturers amend their Instructions for Use accordingly.

Entities:  

Mesh:

Year:  2011        PMID: 22149229     DOI: 10.1583/11-3592.1

Source DB:  PubMed          Journal:  J Endovasc Ther        ISSN: 1526-6028            Impact factor:   3.487


  5 in total

1.  Target vessel displacement during fenestrated and branched endovascular aortic repair and its implications for the role of traditional computed tomography angiography roadmaps.

Authors:  Marloes M Jansen; Merel van der Stelt; Stefan P M Smorenburg; Cornelis H Slump; Joost A van Herwaarden; Constantijn E V B Hazenberg
Journal:  Quant Imaging Med Surg       Date:  2021-09

2.  Quantifying the Functional Stiffness of Pullthrough Wires Used for Endovascular Aneurysm Repairs Using Comparative Tension Dynamometry.

Authors:  Arindam Chaudhuri; Frederic Heim; Nabil Chakfe
Journal:  EJVES Vasc Forum       Date:  2022-05-28

Review 3.  Vascular complications of transcatheter aortic valve replacement: A concise literature review.

Authors:  Muhammad Ali Chaudhry; Muhammad Rizwan Sardar
Journal:  World J Cardiol       Date:  2017-07-26

4.  Are All Wires Created the Same? A Quality Assurance Study of the Stiffness of Wires Typically Employed During Endovascular Surgery Using Tension Dynamometry.

Authors:  Arindam Chaudhuri; Frederic Heim; Nabil Chakfe
Journal:  EJVES Vasc Forum       Date:  2021-06-22

5.  Estimating the "Pull" on a Pullthrough Wire: A Pilot Study.

Authors:  Arindam Chaudhuri; Frederic Heim; Nabil Chakfe
Journal:  EJVES Vasc Forum       Date:  2021-01-13
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.