B Dhanya1, A Suganthi, A K Sen, U Sahoo, A K Seth.
Abstract
A simple, precise and rapid RP-HPLC method was developed for the determination of doxazosin mesylate in pharmaceutical formulations. The method was carried out on a Chromolith RP-C(18) column using a mixture of potassium phosphate buffer and methanol (40:60 v/v) and detection was done at 251 nm. The linearity range was 1-5 μg/ml. The retention time of the drug was 3.8 min. The LOD and LOQ were found to be 0.1 μg/ml and 0.5 μg/ml, respectively.Entities:
Keywords: Doxazosin; RP-HPLC; solid dosage forms
Year: 2011 PMID: 22131637 PMCID: PMC3224404 DOI: 10.4103/0250-474X.89772
Source DB: PubMed Journal: Indian J Pharm Sci ISSN: 0250-474X Impact factor: 0.975
Fig. 1Chromatogram of doxazosin mesylate
Chromatogram of doxazosin mesylate (5 μg/ml) in potassium phosphate buffer and methanol (40:60 v/v)
SUMMARY OF VALIDATION PARAMETERS