D Erkapic1, F Amberger, P Bushoven, J Ehrlich.
Abstract
We report the first clinical case of ineffective high-voltage therapy with 36 J and subsequent effective therapy with 40 J in a patient with electrical storm who had previously received a high-energy implantable cardioverter defibrillator (ICD, Fortify VR, 1233-40Q St. Jude Medical, Sylmar, CA, USA). Using a combination of high energy and optimized fixed millisecond duration biphasic waveform shock, successful defibrillation could be achieved at 8 J below the programmed maximum energy level.Entities:
Mesh:
Year: 2011 PMID: 22124799 DOI: 10.1007/s00399-011-0159-5
Source DB: PubMed Journal: Herzschrittmacherther Elektrophysiol ISSN: 0938-7412