Literature DB >> 22121923

Improved estimation of glomerular filtration rate (GFR) by comparison of eGFRcystatin C and eGFRcreatinine.

Anders Grubb1, Ulf Nyman, Jonas Björk.   

Abstract

OBJECTIVE: GFR-prediction equations based upon cystatin C and creatinine have better diagnostic performance in estimating GFR than equations based upon only one of the two markers. The present work concerns in what way a comparison between separate estimations of GFR based upon cystatin C (eGFR(cystatin C)) or creatinine (eGFR(creatinine)) can be used to evaluate the diagnostic performance of a combined cystatin C- and creatinine-based estimation of GFR.
METHODS: The difference between eGFR(cystatin C) and eGFR(creatinine) was compared with measured GFR (iohexol clearance) and a combined cystatin C- and creatinine-based estimation of GFR in a Swedish-Caucasian cohort of 857 adult patients.
RESULTS: A difference between eGFR(cystatin C) and eGFR(creatinine) of ≥ 40% indicated a markedly reduced diagnostic performance of the combined cystatin C- and creatinine-based estimation of GFR.
CONCLUSION: Comparison of the agreement between eGFR(cystatin C) and eGFR(creatinine) can be used to evaluate the diagnostic performance of combined cystatin C- and creatinine-based estimations of GFR. If 'threshold values' for discordance are exceeded, it must be considered whether the clinical context requires the use of an invasive gold standard method to measure GFR. In some clinical contexts either creatinine or cystatin C are known to be invalidated as markers of GFR and in these situations the use of only the cystatin C- or the creatinine-based GFR estimate should be considered when the 'threshold values' are exceeded.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 22121923      PMCID: PMC3279136          DOI: 10.3109/00365513.2011.634023

Source DB:  PubMed          Journal:  Scand J Clin Lab Invest        ISSN: 0036-5513            Impact factor:   1.713


Introduction

GFR-prediction equations based upon cystatin C (eGFRcystatin C) or creatinine (eGFRcreatinine) may produce estimated GFR-values, of which 80–85% are within ± 30% of GFR measured by invasive gold standard methods. However, the highest percentages of estimated GFR-values within ± 30% of measured GFR are obtained using GFR-prediction equations based upon both cystatin C and creatinine (eGFR cystatin C + creatinine [1-9]. The performance of eGFRcreatinine is reduced inter alia if a patient has an abnormally low or high muscle mass, recently ingested boiled meat or is treated with a drug that influences the tubular secretion of creatinine. The performance of eGFRcystatin C is reduced inter alia if a patient is treated with large doses of glucocorticoids. In such clinical situations the diagnostic performance of a GFR-prediction equation based upon both cystatin C and creatinine may be inferior to those equations based upon only one of the GFR-markers [1]. See also www.egfr.se. However, such situations may not always be recognized by those ordering a GFR-estimate or by the laboratory performing the tests. It has therefore been suggested that a comparison between separate estimations of GFR based upon cystatin C or creatinine can be used to evaluate the diagnostic performance of a combined cystatin C- and creatinine-based estimation of GFR [1]. The present work concerns the relation between the agreement between eGFR cystatin C and eGFR creatinine and the diagnostic performance of eGFR cystatin C + creatinine.

Material and methods

The patient population studied was identical to the one previously used to analyse various equations to combine creatinine and cystatin C to predict GFR [8]. It consisted of adult patients (Swedish-Caucasians ≥ 18 years) consecutively referred to the Departments of Clinical Chemistry, University Hospitals of Lund and Malmö for determination of GFR by iohexol clearance. Simultaneous measurements of plasma creatinine, plasma cystatin C, weight and height were performed and age and gender recorded. The Lund population consisted of 451 patients (225 females) and the Malmö population of 425 patients of whom 19 patients were excluded because of missing plasma creatinine values (n = 6), missing plasma cystatin C values (n = 8) or technical assay errors (n = 5) leaving 406 subjects in the Malmö cohort. All procedures involving subjects and data were in accordance with the Helsinki Declaration of 1975 concerning ethical principles for medical research involving human subjects. The characteristics of the two cohorts and the combined set (n = 857) are shown in Table I and included 12 patients with neurological diseases and secondary muscular atrophy. Common indications for referral were diagnosis and follow-up of chronic kidney disease, evaluation of renal function prior to dosage of drugs cleared by the kidneys, evaluation of potential renal donors, follow-up of unilaterally nephrectomized patients, pre-operative evaluation of patients with hyperparathyroidism and control of renal transplants (n = 44).
Table I

Demographic and anthropometric patient characteristics, plasma creatinine, plasma cystatin C, and iohexol clearance given as median values (2.5 and 97.5 percentiles) in the Lund and Malmö cohorts as well as in the combined set.

ParametersLund (n = 451)Malmö (n = 406)Combined (n = 857)
Age (years)58 (24–83)61 (26–85)59 (26–85)
Females225 (50%)152 (37%)377 (44%)
Total body weight (kg)73 (46–115)78 (49–111)75 (48–112)
Height (cm)170 (151–189)173 (152–190)171 (152–189)
Body surface area (m2)1.83 (1.45–2.31)1.92 (1.49–2.33)1.88 (1.46–2.31)
Body mass index (kg/m2)25 (18–39)26 (18–38)25 (18–38)
Plasma creatinine (μmol/L)92 (39–400)136 (54–623)106 (44–545)
Cystatin C (mg/L)1.18 (0.79–3.07)1.53 (0.85–4.06)1.30 (0.81–3.82)
Iohexol clearance (mL/min per 1.73 m2)63 (11–124)42 (8–115)55 (9–121)
Demographic and anthropometric patient characteristics, plasma creatinine, plasma cystatin C, and iohexol clearance given as median values (2.5 and 97.5 percentiles) in the Lund and Malmö cohorts as well as in the combined set.

Determination of iohexol clearance

Five mL of iohexol (Omnipaque 300 mg iodine/mL, GE Healthcare, Oslo, Norway) were administered intravenously in an antecubital vein. Iohexol clearance (referred to as ‘measured GFR') was calculated from plasma clearance of a single plasma sample of iohexol [10] drawn at varying times, normally 4 hours after injection, according to expected GFR as determined by plasma creatinine concentration and anthropometric data. The exact time of administration and blood sampling were documented by a specialist nurse. Plasma iohexol concentrations were determined by high-pressure liquid chromatography with a total analytical variation of 2–4% (coefficient of variation, CV%) at the range of iohexol concentrations normally encountered during the study [11]. The Dubois formula was used to adjust the measured GFR values to 1.73 m2 body surface area [12].

Determination of plasma creatinine

Plasma concentrations of creatinine were determined at Lund University Hospital by an enzymatic colorimetric assay on a Hitachi Modular P analyzer (Roche Diagnostics, Mannheim, Germany) and with a calibrator traceable to primary reference material with values assigned by isotope dilution mass spectrometry (IDMS) [13]. At Malmö University Hospital a modified Jaffe colorimetric method was used on a Beckman LX20 analyzer (Beckman Coulter, Inc., Fullerton, CA, USA) employing zero-point calibration and a calibrator traceable to primary reference material with values assigned by IDMS [14,15]. Total analytical variation (CV%) of the enzymatic method in Lund was 1.4–3.0% at concentrations of creatinine between 60 and 578 μmol/L and 2.2–2.8% at concentrations between 53 and 631 μmol/L for the Jaffe method in Malmö.

Determination of plasma cystatin C

Plasma cystatin C levels were determined by an automated particle-enhanced immunoturbidimetric method [16] using a Hitachi Modular P analysis system, reagents (code Nos LX002, S2361, X0973, X0974) obtained from DakoCytomation (Glostrup, Denmark) and following the procedure recommended by the reagent producer. The procedure had a total coefficient of variation of 2.1% at a cystatin C level of 1.0 mg/L and of 1.7% at a level of 4.0 mg/L. All samples were analysed within one day after collection or frozen at −20°C until analysed.

Prediction equations

The Lund-Malmö creatinine-based equation (eGFRcreatinine) with age and gender [17] and the Grubb cystatin C-based equation (eGFRcystatin C) based on adults and including gender [2] were selected for the present analysis. Plasma creatinine (pCr) is expressed in μmol/L, plasma cystatin C (pCy) in mg/L, age in years and ln denotes the natural logarithm. Both equations express relative GFR in mL/min per 1.73 m2 body surface area.

Lund-Malmö creatinine equation (eGFRcreatinine)

GFR = eX − 0.0124 × age + 0.339 × ln(age) − 0.226 (if female) X = 4.62 − 0.0112 × pCr (if pCr < 150 μmol/L) X = 8.17 + 0.0005 × pCr − 1.07 × ln(pCr) (if pCr ≥ 150 μmol/L)

Grubb cystatin C equation (eGFRcystatin C)

GFR = 86.49 × pCy−1.686 × 0.948 (if female), equivalent with GFR = e4.46 − 1.686 × ln (pCy) − 0.053 (if female) GFR estimates from the combined use of the two analytes (eGFRcystatin C + creatinine) were based on the arithmetic mean of eGFRcystatin C and eGFRcreatinine, which has proved as accurate as more complex equations [8].

Statistical evaluation

All statistical analyses were conducted using SPSS release 18.0.1. (SPSS Inc, Chicago, USA). In the statistical testing we regarded p-values in the order of 0.05 as moderate evidence against the null hypothesis, whereas p-values in the order of 0.001 or below were regarded as strong evidence against the null hypothesis [18]. The present study focused on the accuracy of the arithmetic mean of eGFRcystatin C and eGFR creatinine, denoted eGFRcystatin C + creatinine, in relation to the agreement between eGFRcystatin C and eGFRcreatinine. The accuracy of eGFRcystatin C + creatinine was reflected by the absolute percentage error: |eGFRcystatin C + creatinine − measured GFR| /measured GFR, and summarized as the percentage of estimates within 30% (P30) and 10% (P10) of measured GFR [19]. The agreement was reflected by the difference%, i.e. the absolute difference |eGFRcystatin CeGFRcreatinine| expressed in percent relative to the arithmetic mean eGFR cystatin C + creatinine. The following analyses were made: Pearson's and Spearman's correlation coefficients (denoted r and rs) were used to evaluate the overall association between accuracy (absolute percentage error) and agreement (difference%). The accuracy categorized as P30 and P10 was evaluated in relation to agreement (difference%) rounded to nearest integer and then categorized as < 10%, 10–19%, 20–29%, 30–39% and ≥40% difference. Fisher's exact test was used to evaluate differences in P30 and P10 across categories of agreement. Measured GFR is related to both accuracy and agreement and may thus confound the association between agreement and accuracy. To account for such confounding we modelled accuracy, i.e. P30 and P10, respectively, using logistic regression with measured GFR and difference% as continuous covariates. We calculated an ‘improvement index', defined as the proportion of all GFR estimates for which eGFRcystatin C + creatinine, but not both eGFRcystatin C and eGFRcreatinin, were inaccurate according to P30 and P10, respectively. This index represents the upper limit of the improvement in accuracy that could be obtained if the most accurate eGFR (i.e. eGFR cystatin C + creatinine, eGFR cystatin C or eGFR creatinin) was consistently applied for each patient. Note that the sum of, e.g. P30 and the corresponding improvement index for P30 can never exceed 100%. The improvement index depends on the accuracy, i.e. the potential for improvement is higher when accuracy is low. To account for such confounding in the association between agreement and the improvement index, we modelled the improvement index using logistic regression with inaccuracy (absolute percentage error) and difference% as continuous covariates.

Results

The overall association between difference% and absolute percentage error was not consistent (r = 0.13, p<0.001 but rs = 0.05, p = 0.12), however, P30 was clearly decreased for differences between eGFRcystatin C and eGFRcreatinin exceeding a ‘threshold value’ of 40% (Table II; p = 0.02 when comparing P30 for 30–39 and >40% difference). The dip in accuracy when expressed as P10 seemed to occur already at 30–39% difference, but the statistical evidence for this dip was weak (p = 0.09 when comparing P10 for 20–29 and 30–39% difference). Measured GFR was noticeably lower for differences exceeding 40%, but the suggested inverse association between difference% and P30 remained clear when measured GFR was adjusted for using logistic regression (p = 0.001), whereas the association between difference% and P10 remained weaker (p = 0.05).
Table II

Accuracy of eGFR cystatin C + creatinine, the arithmetic mean of eGFRcystatin C and eGFRcretinine, calculated as the percentage of estimates within 30% (P30) and 10% (P10) of measured GFR in relation to the difference% between eGFRcystatin C and eGFRcreatinine, defined as |eGFRcystatin C − eGFRcreatinine| / eGFRcystatin C + creatinine. The improvement index, defined as the proportion of all GFR estimates where eGFRcystatin C + creatinine, but not both eGFRcystatin C and creatinine, were inaccurate within 30% and 10% of measured GFR, is also presented.

Accuracy (%)Improvement index (%)
Difference %Median measured GFR (mL/min per 1.73 m2)30%10%30%10%
<10 (n = 220)5990.043.61.88.2
10–19 (n = 200)5991.548.52.525.0
20–29 (n = 175)6194.345.15.739.4
30–39 (n = 120)5490.035.08.352.5
≥ 40 (n = 142)3979.640.818.330.3
Total (n = 857)5589.543.46.428.4
Accuracy of eGFR cystatin C + creatinine, the arithmetic mean of eGFRcystatin C and eGFRcretinine, calculated as the percentage of estimates within 30% (P30) and 10% (P10) of measured GFR in relation to the difference% between eGFRcystatin C and eGFRcreatinine, defined as |eGFRcystatin CeGFRcreatinine| / eGFRcystatin C + creatinine. The improvement index, defined as the proportion of all GFR estimates where eGFRcystatin C + creatinine, but not both eGFRcystatin C and creatinine, were inaccurate within 30% and 10% of measured GFR, is also presented. The improvement index generally suggested higher potential for improvement in accuracy when the difference between eGFRcystatin C and eGFRcreatinin was considerable (Table II). This association between the agreement (difference%) and the potential for improvement in accuracy remained evident when differences in accuracy across levels of agreement were adjusted for using logistic regression (p<0.001 both for improvement in P30 and in P10).

Discussion

The diagnostic performance of eGFRcreatinine is reduced inter alia if a patient has an abnormally low or high muscle mass, recently ingested boiled meat or is treated with a drug that influences the tubular secretion of creatinine. In these clinical contexts the diagnostic performance of eGFRcystatin C is generally unaltered. However, the performance of eGFR cystatin C is impaired if a patient is treated with large doses of glucocorticoids and in this situation the performance of eGFRcreatinine is still acceptable. Although GFR-prediction equations based upon both cystatin C and creatinine (eGFRcystatin C + creatinine) generally are superior to GFR-prediction equations based upon either cystatin C (eGFRcystatin C) or creatinine (eGFRcreatinine) this may not be the case in these specific clinical contexts. Although these contexts may be easily recognized in some cases, they will not invariably be recognized. There may also be additional, not yet identified, clinical contexts invalidating either cystatin C or creatinine as useful markers for GFR. Comparing eGFR cystatin C and eGFRcreatinine might be helpful to identify both known and unknown causes when neither cystatin C nor creatinine are suitable as a marker for GFR [1]. To be able to efficiently use such a comparison, it must be known when the discordance between eGFRcystatin C and eGFRcreatinine is large enough to indicate such a condition. The present study based upon measured GFR and eGFRcystatin C and eGFRcreatinine in a patient cohort of 857 Swedish-Caucasian adult patients indicates that if the discordance is 40% or more, the diagnostic performance of eGFRcystatin C + creatinine is markedly reduced. Such discordance should initiate a more careful evaluation of the clinical context to disclose conditions invalidating either creatinine or cystatin C as a GFR marker. If such conditions are identified, GFR might be best estimated using a prediction equation based upon only the non-invalidated marker. If such conditions are not identified, it should be realized that the estimation of GFR is unreliable and that an invasive, gold standard, measurement of GFR might be required. It should be realized, that the discordance value, the ‘threshold value', indicating requirement of a further evaluation of the clinical context to improve estimation of GFR, as presented in this work, is influenced by the actual patient cohort and by the equations used for estimating GFR, i.e. eGFR cystatin C, eGFRcreatinine and eGFRcystatin C + creatinine. For other patient cohorts might contain proportionally more, or fewer patients, with conditions invalidating either cystatin C or creatinine as a GFR marker. The more patients with such conditions in the cohort, the greater the potential for improvement of eGFR by comparison of eGFRcystatin C and eGFRcreatinine. The equations for eGFRcystatin C, eGFRcreatinine and eGFRcystatin C + creatinine will also influence the ‘threshold value’ by being more or less sensitive for patient characteristics reducing the value of creatinine and cystatin C as markers for GFR. It is possible to calculate an ‘improvement index', defined as the proportion of all GFR estimates for which eGFRcystatin C + creatinine, but not both eGFRcystatin C and eGFRcreatinin, are inaccurate according to P30 or P10. This index will represent the upper limit of improvement in accuracy that could be obtained, if the most accurate eGFR (i.e. eGFRcystatin C + creatinine, eGFRcystatin C or eGFRcreatinin) was consistently applied. In the present study the improvement index was 18.3 % for a discordance threshold of ≥40%, which means that if an accurate evaluation of the relevant clinical conditions could be performed for each patient the P30-value of 79.6% for eGFRcystatin C + creatinine, would theoretically increase to 79.6 + 18.3 = 97.9%. This ‘improvement index’ will, exactly like the ‘threshold value’ of discordance, also be influenced by the actual patient cohort and by the equations used for eGFR, i.e. eGFRcystatin C, eGFRcreatinine and eGFRcystatin C + creatinine and for the same reasons.
  19 in total

Review 1.  Sifting the evidence-what's wrong with significance tests?

Authors:  J A Sterne; G Davey Smith
Journal:  BMJ       Date:  2001-01-27

2.  Creatininium reference intervals for corrected methods.

Authors:  A Mårtensson; P Rustad; H Lund; H Ossowicki
Journal:  Scand J Clin Lab Invest       Date:  2004       Impact factor: 1.713

3.  A formula to estimate the approximate surface area if height and weight be known. 1916.

Authors:  D Du Bois; E F Du Bois
Journal:  Nutrition       Date:  1989 Sep-Oct       Impact factor: 4.008

4.  Simple cystatin C-based prediction equations for glomerular filtration rate compared with the modification of diet in renal disease prediction equation for adults and the Schwartz and the Counahan-Barratt prediction equations for children.

Authors:  Anders Grubb; Ulf Nyman; Jonas Björk; Veronica Lindström; Bengt Rippe; Gunnar Sterner; Anders Christensson
Journal:  Clin Chem       Date:  2005-06-16       Impact factor: 8.327

5.  GFR is better estimated by considering both serum cystatin C and creatinine levels.

Authors:  Yann Bouvet; François Bouissou; Yvon Coulais; Sophie Séronie-Vivien; Mathieu Tafani; Stéphane Decramer; Etienne Chatelut
Journal:  Pediatr Nephrol       Date:  2006-06-22       Impact factor: 3.714

6.  Serum cystatin C, determined by a rapid, automated particle-enhanced turbidimetric method, is a better marker than serum creatinine for glomerular filtration rate.

Authors:  J Kyhse-Andersen; C Schmidt; G Nordin; B Andersson; P Nilsson-Ehle; V Lindström; A Grubb
Journal:  Clin Chem       Date:  1994-10       Impact factor: 8.327

7.  Glomerular filtration rate estimated by cystatin C among different clinical presentations.

Authors:  A D Rule; E J Bergstralh; J M Slezak; J Bergert; T S Larson
Journal:  Kidney Int       Date:  2006-01       Impact factor: 10.612

8.  Plasma clearance of a new contrast agent, iohexol: a method for the assessment of glomerular filtration rate.

Authors:  E Krutzén; S E Bäck; I Nilsson-Ehle; P Nilsson-Ehle
Journal:  J Lab Clin Med       Date:  1984-12

9.  A method for the calculation of renal clearance based on a single plasma sample.

Authors:  L Jacobsson
Journal:  Clin Physiol       Date:  1983-08

Review 10.  Non-invasive estimation of glomerular filtration rate (GFR). The Lund model: Simultaneous use of cystatin C- and creatinine-based GFR-prediction equations, clinical data and an internal quality check.

Authors:  Anders Grubb
Journal:  Scand J Clin Lab Invest       Date:  2010-04       Impact factor: 1.713

View more
  21 in total

1.  Estimating glomerular filtration rate for the full age spectrum from serum creatinine and cystatin C.

Authors:  Hans Pottel; Pierre Delanaye; Elke Schaeffner; Laurence Dubourg; Bjørn Odvar Eriksen; Toralf Melsom; Edmund J Lamb; Andrew D Rule; Stephen T Turner; Richard J Glassock; Vandréa De Souza; Luciano Selistre; Karolien Goffin; Steven Pauwels; Christophe Mariat; Martin Flamant; Natalie Ebert
Journal:  Nephrol Dial Transplant       Date:  2017-03-01       Impact factor: 5.992

2.  Estimating glomerular filtration rate in children: evaluation of creatinine- and cystatin C-based equations.

Authors:  Cathrin L Salvador; Camilla Tøndel; Alexander D Rowe; Anna Bjerre; Atle Brun; Damien Brackman; Lars Mørkrid
Journal:  Pediatr Nephrol       Date:  2018-08-31       Impact factor: 3.714

3.  Measuring and estimating glomerular filtration rate in children.

Authors:  Hans Pottel
Journal:  Pediatr Nephrol       Date:  2016-04-26       Impact factor: 3.714

Review 4.  Estimating GFR prior to contrast medium examinations--what the radiologist needs to know!

Authors:  Ulf Nyman; Jonas Björk; Sten-Erik Bäck; Gunnar Sterner; Anders Grubb
Journal:  Eur Radiol       Date:  2015-05-28       Impact factor: 5.315

5.  Comparison of Creatinine and Cystatin C to Estimate Renal Function in Geriatric and Frail Patients.

Authors:  Erik Dahlén; Linda Björkhem-Bergman
Journal:  Life (Basel)       Date:  2022-06-07

Review 6.  Update on cystatin C: incorporation into clinical practice.

Authors:  Michael G Shlipak; Monica D Mattes; Carmen A Peralta
Journal:  Am J Kidney Dis       Date:  2013-05-20       Impact factor: 8.860

7.  Performance of the creatinine and cystatin C-based equations for estimation of GFR in Chinese patients with chronic kidney disease.

Authors:  Min Yang; Guang Xu; Lilu Ling; Jianying Niu; Tong Lu; Xin Du; Yong Gu
Journal:  Clin Exp Nephrol       Date:  2016-04-28       Impact factor: 2.801

8.  Combining GFR estimates from cystatin C and creatinine-what is the optimal mix?

Authors:  Emil den Bakker; Reinoud Gemke; Joanna A E van Wijk; Isabelle Hubeek; Birgit Stoffel-Wagner; Arend Bökenkamp
Journal:  Pediatr Nephrol       Date:  2018-05-17       Impact factor: 3.714

9.  Association of a cystatin C gene variant with cystatin C levels, CKD, and risk of incident cardiovascular disease and mortality.

Authors:  Conall M O'Seaghdha; Adrienne Tin; Qiong Yang; Ronit Katz; Yongmei Liu; Tamara Harris; Brad Astor; Josef Coresh; Caroline S Fox; W H Linda Kao; Michael G Shlipak
Journal:  Am J Kidney Dis       Date:  2013-08-07       Impact factor: 8.860

10.  Impact of creatinine production on the agreement between glomerular filtration rate estimates using cystatin C-derived, and 4- and 6-variable Modification of Diet in Renal Disease (MDRD), and Chronic Kidney Disease Epidemiology Collaboration (CKD-EPI) equations.

Authors:  Esperanza F Hermida-Cadahia; Natalia Lampon; J Carlos Tutor
Journal:  Ups J Med Sci       Date:  2012-07-03       Impact factor: 2.384

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.