J C Chu1, L J Solin, C C Hwang, B Fowble, G E Hanks, R L Goodman.
Abstract
Effective radiation therapy to intact breasts requires the delivery of adequate dose to a large target volume using complex beam arrangements. A semi-empirical method is described to determine the correct gantry, collimator, and couch positions for a geometrically accurate field match among adjoining radiation beams. The technique uses a metal rod and chain combination to aid determination of the proper couch setting under remote fluoroscopy control. A couch position error of more than half a degree is easily detectable by this technique.Mesh:
Year: 1990 PMID: 2211240 DOI: 10.1016/0360-3016(90)90030-n
Source DB: PubMed Journal: Int J Radiat Oncol Biol Phys ISSN: 0360-3016 Impact factor: 7.038