Literature DB >> 22109977

Duration of balloon inflation for optimal stent deployment: five seconds is not enough.

Thomas Hovasse1, Darren Mylotte, Philippe Garot, Neus Salvatella, Marie-Claude Morice, Bernard Chevalier, Augusto Pichard, Thierry Lefèvre.   

Abstract

OBJECTIVES: To assess the effect of the duration of stent inflation on stent expansion using digital stent enhancement (DSE).
BACKGROUND: Optimal stent expansion and apposition to the vessel wall are of critical importance to optimize the results of percutaneous coronary intervention (PCI). However, it is not known if stent inflation duration impacts on stent expansion.
METHODS: We performed a prospective cohort study in patients undergoing PCI. Quantitative coronary angiography and DSE data were analyzed. DSE was performed at 5, 15, and 25 sec during stent implantation, after target balloon inflation pressure was achieved.
RESULTS: One hundred and four consecutive patients (150 lesions) were enrolled. The mean age was 66.9 ± 11.1 years. Complex lesions (ACC/AHA B2/C) occurred in 26.9%. Stents used: Cypher Select (54.1%), Xience V (30.6%), and Taxus Liberté (15.3%). The minimal stent diameter increased significantly with the duration of stent inflation: 2.60 ± 0.51, 2.76 ± 0.51, and 2.82 ± 0.52 mm at 5, 15, and 25 sec (P < 0.0001). Similarly, maximal stent diameter increased with the duration of stent inflation: 3.21 ± 0.51, 3.32 ± 0.52, and 3.36 ± 0.54 mm (P < 0.0001). The average stent diameter also increased with longer stent inflation (P < 0.0001). Using MUSIC criteria 24.0, 53.3, and 68.0% of stents were appropriately expanded at 5, 15, and 25 sec (P < 0.0001).
CONCLUSIONS: The duration of stent balloon inflation has a significant impact on stent expansion. Stent deployment for >25 sec is recommended.
Copyright © 2012 Wiley Periodicals, Inc.

Entities:  

Mesh:

Year:  2012        PMID: 22109977     DOI: 10.1002/ccd.23343

Source DB:  PubMed          Journal:  Catheter Cardiovasc Interv        ISSN: 1522-1946            Impact factor:   2.692


  5 in total

Review 1.  Hemodynamic Support Devices for Shock and High-Risk PCI: When and Which One.

Authors:  George W Vetrovec
Journal:  Curr Cardiol Rep       Date:  2017-08-31       Impact factor: 2.931

2.  Effect of stent inflation pressure and post-dilatation on the outcome of coronary artery intervention. A report of more than 90,000 stent implantations.

Authors:  Ole Fröbert; Giovanna Sarno; Stefan K James; Nawsad Saleh; Bo Lagerqvist
Journal:  PLoS One       Date:  2013-02-13       Impact factor: 3.240

3.  Impact of the Balloon Inflation Time and Pattern on the Coronary Stent Expansion.

Authors:  Jarosław Skowroński; Rafał Wolny; Jan Jastrzębski; Paweł Tyczyński; Karol Szlazak; Jerzy Pręgowski; Gary S Mintz; Karolina Liżewska; Wojciech Świeszkowski; Zbigniew Chmielak; Adam Witkowski
Journal:  J Interv Cardiol       Date:  2019-03-21       Impact factor: 2.279

Review 4.  Prolonged versus brief balloon inflation during arterial angioplasty for de novo atherosclerotic disease: a systematic review and meta-analysis.

Authors:  Mark Rockley; Prasad Jetty; Aleksandar Radonjic; Kathleen Rockley; George Wells; Dean Fergusson
Journal:  CVIR Endovasc       Date:  2019-08-17

5.  A randomized controlled clinical trial of prolonged balloon inflation during stent deployment strategy in primary percutaneous coronary intervention for ST-segment elevation myocardial infarction: a pilot study.

Authors:  Min Ma; Ling Wang; Kai-Yue Diao; Shi-Chu Liang; Ye Zhu; Hua Wang; Mian Wang; Li Zhang; Zhi-Gang Yang; Yong He
Journal:  BMC Cardiovasc Disord       Date:  2022-02-04       Impact factor: 2.298