Image-quality perception as a function of dose in digital radiography.

Thomas Lehnert1, Nagy N N Naguib, Huedayi Korkusuz, Ralf W Bauer, J Matthias Kerl, Martin G Mack, Thomas J Vogl.   

Abstract

OBJECTIVE: The purpose of this article is to determine the degree to which the skin entrance dose could be lowered, by adjusting exposure parameters and filtration, and the subsequent effect on readers' confidence levels of digital radiographs.
MATERIALS AND METHODS: The study was prospectively performed on a cadaver. Digital radiographs of bones were evaluated and scored on a 9-point scale separately by four radiologists who were blinded to the types of filtration and doses used. The study entailed three phases: phase 1, random dose and filter; phase 2, fixed filter and varying radiation doses (100%, 75%, 50%, and 25% of the standard recommended dose); and phase 3, fixed dose and varying filtration (no filtration, aluminum filter, and aluminum-copper filter). Skin entrance dose was measured using a dosimeter placed on the skin. Differences in scores were tested using a Friedman test.
RESULTS: The mean scores given to images with 100%, 75%, 50%, and 25% of the recommended standard dose were 6.18, 6.1, 5.11, and 4.07, respectively. No significant difference was noted between 100%- and 75%-dose images (p = 0.1). A significant difference (p < 0.0001) was noted when we compared the 100%- and 75%-dose images with the 50%- and 25%-dose images. The mean scores given for no filtration, aluminum filtration, and aluminum-copper filtration were 5.67, 5.43, and 5.18, respectively. No significant difference between no filtration and aluminum filtration (p = 0.411) was noted. A significant difference was detected between no filtration and aluminum-copper filtration (p = 0.012). The combination of an aluminum filter and a 75% standard dose achieved a 31.1% reduction in skin entrance dose.
CONCLUSION: It is possible to achieve a 31.1% reduction in skin entrance dose for imaging bony structures by using 75% of the standard dose and aluminum filtration without significantly affecting image quality.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 22109295     DOI: 10.2214/AJR.10.6269

Source DB:  PubMed          Journal:  AJR Am J Roentgenol        ISSN: 0361-803X            Impact factor:   3.959


  2 in total

1.  [Dose reduction and adequate image quality in digital radiography: a contradiction?].

Authors:  S Pötter-Lang; M Dünkelmeyer; M Uffmann
Journal:  Radiologe       Date:  2012-10       Impact factor: 0.635

2.  Development and validation of a visual grading scale for assessing image quality of AP pelvis radiographic images.

Authors:  Hussien A A B Mraity; Andrew England; Simon Cassidy; Peter Eachus; Alejandro Dominguez; Peter Hogg
Journal:  Br J Radiol       Date:  2016-03-04       Impact factor: 3.039

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.