The reliability and accuracy of a noncontact electrocardiograph system for screening purposes.

Michael Czaplik1, Benjamin Eilebrecht, Rafael Walocha, Marian Walter, Patrick Schauerte, Steffen Leonhardt, Rolf Rossaint.   

Abstract

BACKGROUND: Electrocardiography (ECG) requires the application of electrodes to the skin and often necessitates undressing. Capacitively coupled electrodes embedded in a normal chair would be a rational alternative for ECG screening. We evaluated the reliability and accuracy of ECG electrodes imbedded in a chair cushion.
METHODS: Two independent clinicians compared ECG recordings obtained using skin electrodes with recordings obtained using capacitively coupled electrodes that were embedded in a chair cushion in an anesthesiology premedication room, a cardiology outpatient ward, and a cardiology day ward. We analyzed the data to compare the sensitivity and specificity for the diagnosis of cardiac arrhythmias.
RESULTS: ECG recordings were obtained from 107 patients. Heart rate was accurately measured using the capacitively coupled electrodes, but motion artifacts made the identification of P and T waves unreliable. Signal quality was poor for patients with low body weight, patients wearing clothing containing mixed fibers, and patients wearing sweaty shirts.
CONCLUSIONS: Heart rate was accurately measured, and some cardiac arrhythmias were correctly diagnosed using capacitive ECG electrodes. Capacitive electrodes embedded into an examination chair are a promising tool for preoperative screening. Improved artifact reduction algorithms are needed before capacitive electrodes will replace skin electrodes.

Entities:  

Mesh:

Year:  2011        PMID: 22104072     DOI: 10.1213/ANE.0b013e31823f4182

Source DB:  PubMed          Journal:  Anesth Analg        ISSN: 0003-2999            Impact factor:   5.108


  2 in total

1.  UnoViS: the MedIT public unobtrusive vital signs database.

Authors:  Tobias Wartzek; Michael Czaplik; Christoph Hoog Antink; Benjamin Eilebrecht; Rafael Walocha; Steffen Leonhardt
Journal:  Health Inf Sci Syst       Date:  2015-06-02

2.  Reduction of Artifacts in Capacitive Electrocardiogram Signals of Driving Subjects.

Authors:  Tamara Škorić
Journal:  Entropy (Basel)       Date:  2021-12-22       Impact factor: 2.524

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.