Literature DB >> 22099893

Evaluation of accuracy, reliability, and repeatability of five dental pulp tests.

Eugene Chen1, Paul V Abbott.   

Abstract

INTRODUCTION: The aim of this study was to compare the clinical accuracy, reliability, and repeatability of laser Doppler flowmetry (LDF), an electric pulp test (EPT), and various thermal pulp sensibility tests.
METHODS: Pulp tests were done on 121 teeth in 20 subjects by using LDF, EPT, and thermal pulp testing (CO(2), Endo Frost [EF], Ice) during 2 or 3 test sessions with at least 1-week intervals. The order of testing was reversed on the second visit. A laser Doppler flowmeter was used to measure mean pulp blood flow (Flux) calibrated against a brownian motion medium and zeroed against a static reflector. The laser source was 780 nm, with 0.5-mm fiber separation in the probe, 3.1 kHz as the primary bandwidth for filter set to 0.1-second time output constant. Customized polyvinylsiloxane splints were fabricated for each participant, and a minimum of 90-second recording time was used for each tooth. Raw data were analyzed by using repeated measure analysis of variance, pairwise comparisons, and interclass correlations (ICC).
RESULTS: The accuracy of EPT, CO(2), and LDF tests was 97.7%, 97.0%, and 96.3%, respectively, without significant differences (P > .3). Accuracy of EF and Ice was 90.7% and 84.8%, respectively. EPT (P = .015) and CO(2) (P = .022) were significantly more accurate than EF. LDF was more accurate than EF, but this was not statistically significant (P = .063). Ice was significantly less accurate than EPT (P = .004), CO(2) (P = .005), LDF (P = .006), and EF (P = .019). With the exception of Ice (effect of visit: F(2,38) = 5.67, mean squared error = 0.01, P = .007, η(2)(p) = 0.23), all tests were reliable. Ice (ICC = 0.677) and LDF (ICC = 0.654) were the most repeatable of the tests, whereas EPT (ICC = 0.434) and CO(2) (ICC = 0.432) were less repeatable.
CONCLUSIONS: CO(2), EPT, and LDF were reliable and the most accurate tests, but CO(2) and EPT were less repeatable yet less time-consuming than LDF. EF was reliable but not as accurate as EPT and CO(2) and less repeatable than Ice and LDF. Ice was the most repeatable but the least accurate and least reliable test.
Copyright © 2011 American Association of Endodontists. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 22099893     DOI: 10.1016/j.joen.2011.07.004

Source DB:  PubMed          Journal:  J Endod        ISSN: 0099-2399            Impact factor:   4.171


  13 in total

1.  Effect of extracorporeal shockwave therapy (ESWT) on pulpal blood flow after orthodontic treatment: a randomized clinical trial.

Authors:  Frank Falkensammer; Wolfgang Schaden; Christoph Krall; Josef Freudenthaler; Hans-Peter Bantleon
Journal:  Clin Oral Investig       Date:  2015-07-17       Impact factor: 3.573

2.  How standard deviation contributes to the validity of a LDF signal: a cohort study of 8 years of dental trauma.

Authors:  Herman J J Roeykens; Peter De Coster; Wolfgang Jacquet; Roeland J G De Moor
Journal:  Lasers Med Sci       Date:  2019-05-16       Impact factor: 3.161

3.  Evaluation of the Response to Pulpal Sensibility Tests (Cold, EPT) in Anemic and Healthy Women.

Authors:  Maryam Kazemipoor; Mohammad Kazem Vatanchian; Sara Jambarsang; Fatemeh Owlia
Journal:  Int J Dent       Date:  2022-06-18

Review 4.  Use of preoperative cone-beam computed tomography to aid in establishment of endodontic working length: A systematic review and meta-analysis.

Authors:  Andrew Paterson; Vittorio Franco; Shanon Patel; Federico Foschi
Journal:  Imaging Sci Dent       Date:  2020-09-16

Review 5.  A Diagnostic Insight of Dental Pulp Testing Methods in Pediatric Dentistry.

Authors:  Andreea Igna; Doina Mircioagă; Marius Boariu; Ștefan-Ioan Stratul
Journal:  Medicina (Kaunas)       Date:  2022-05-16       Impact factor: 2.948

6.  Correlation between Histological Status of the Pulp and Its Response to Sensibility Tests.

Authors:  Mandana Naseri; Akbar Khayat; Sara Zamaheni; Shiva Shojaeian
Journal:  Iran Endod J       Date:  2017

7.  Diagnostic accuracy of three placement sites for the cold test in subjects amongst different age groups.

Authors:  Brenda Eréndida Castillo-Silva; Jorge Alejandro Alegría-Torres; Gabriel Alejandro Martínez-Castañón; Carlo Eduardo Medina-Solís; Norma Verónica Zavala-Alonso; Nereyda Niño-Martínez; Eva Concepción Aguirre-López; Nuria Patiño-Marín
Journal:  BMC Oral Health       Date:  2019-08-19       Impact factor: 2.757

8.  Relationship between the difference in electric pulp test values and the diagnostic type of pulpitis.

Authors:  Huachao Sui; Yangyang Lv; Mo Xiao; Liwen Zhou; Feng Qiao; Jinxin Zheng; Cuicui Sun; Jieni Fu; Yufan Chen; Yimeng Liu; Jie Zhou; Ligeng Wu
Journal:  BMC Oral Health       Date:  2021-07-10       Impact factor: 2.757

9.  Partial Necrosis Consequence of the Infection Spreading from an Adjacent Apical Periodontitis: A Case Report.

Authors:  Saeed Asgary; Leyla Roghanizadeh
Journal:  Iran Endod J       Date:  2018

10.  Levels of matrix metalloproteinase-8 and cold test in reversible and irreversible pulpitis.

Authors:  Eva Concepción Aguirre-López; Nuria Patiño-Marín; Gabriel Alejandro Martínez-Castañón; Carlo Eduardo Medina-Solís; Brenda Eréndida Castillo-Silva; Oscar Cepeda-Argüelles; Luis Alejandro Aguilera-Galaviz; Pedro Rosales-García
Journal:  Medicine (Baltimore)       Date:  2020-12-24       Impact factor: 1.817

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.