Evaluating the material parameters of the human cornea in a numerical model.

Wiesław Sródka1.   

Abstract

PURPOSE: The values of the biomechanical human eyeball model parameters reported in the literature are still being disputed. The primary motivation behind this work was to predict the material parameters of the cornea through numerical simulations and to assess the applicability of the ubiquitously accepted law of applanation tonometry - the Imbert-Fick equation.
METHODS: Numerical simulations of a few states of eyeball loading were run to determine the stroma material parameters. In the computations, the elasticity moduli of the material were related to the stress sign, instead of the orientation in space.
RESULTS: Stroma elasticity secant modulus E was predicted to be close to 0.3 MPa. The numerically simulated applanation tonometer readings for the cornea with the calibration dimensions were found to be lower by 11 mmHg then IOP = 48 mmHg.
CONCLUSIONS: This discrepancy is the result of a strictly mechanical phenomenon taking place in the tensioned and simultaneously flattened corneal shell and is not related to the tonometer measuring accuracy. The observed deviation has not been amenable to any GAT corrections, contradicting the Imbert-Fick law. This means a new approach to the calculation of corrections for GAT readings is needed.

Entities:  

Mesh:

Year:  2011        PMID: 22098299

Source DB:  PubMed          Journal:  Acta Bioeng Biomech        ISSN: 1509-409X            Impact factor:   1.073


  4 in total

1.  Automatic method of analysis and measurement of additional parameters of corneal deformation in the Corvis tonometer.

Authors:  Robert Koprowski
Journal:  Biomed Eng Online       Date:  2014-11-19       Impact factor: 2.819

2.  Imaging shear stress distribution and evaluating the stress concentration factor of the human eye.

Authors:  S Joseph Antony
Journal:  Sci Rep       Date:  2015-03-10       Impact factor: 4.379

3.  Transient viscous response of the human cornea probed with the Surface Force Apparatus.

Authors:  Bruno Zappone; Navinkumar J Patil; Marco Lombardo; Giuseppe Lombardo
Journal:  PLoS One       Date:  2018-05-25       Impact factor: 3.240

4.  Effects of the LASIK flap thickness on corneal biomechanical behavior: a finite element analysis.

Authors:  Lihua Fang; Yan Wang; Ruizhi Yang; Sijing Deng; Jiahao Deng; Linsun Wan
Journal:  BMC Ophthalmol       Date:  2020-02-24       Impact factor: 2.209

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.