Biomechanical comparison of an interspinous device and a rigid stabilization on lumbar adjacent segment range of motion.

F Hartmann1, S O Dietz, S Kuhn, H Hely, P M Rommens, E Gercek.   

Abstract

PURPOSE OF THE STUDY: Decompression surgery with or without fusion is the gold standard treatment of lumbar spinal stenosis, but adjacent segment degeneration has been reported as a long-term complication after fusion. This led to the development of dynamic implants like the interspinous devices. They are supposed to limit extension and expand the spinal canal at the symptomatic level, but with reduced effect on the range of motion of the adjacent segments. The aim of the present study is the evaluation of the biomechanical effects on the range of motion (ROM) of adjacent lumbar segments after decompression and instrumentation with an interspinous device compared to a rigid posterior stabilization device.
MATERIALS AND METHODS: Eight fresh frozen human cadaver lumbar spines (L2-L5) were tested in a spinal testing device with a moment of 7.5 Nm in flexion/extension, lateral bending and rotation with and without a preload. The preload was applied as a follower load of 400N along the curvature of the spine. The range of motion (ROM) of the adjacent segments L2/L3 and L4/L5 was measured with the intact segment L3/L4, after decompression, consisting of resection of the interspinous ligament, flavectomy and bilateral medial facetecomy, and insertion of the Coflex® (Paradigm Spine, Wurmlingen) and after instrumentation with Click X® (Synthes, Umkirch) as well.
RESULTS: The interspinous and the rigid device caused a significant increase of ROM at both adjacent segments during all directions of motion and under follower load, without significant difference between these devices. The ROM of L2/L3 tends to increase more than the ROM of L4/L5 after instrumentation without statistical significance. DISCUSSION: The "dynamic" Coflex device caused a significant increase of ROM at both adjacent lumbar segments comparable to the increase of ROM after instrumentation with the rigid Click X device. Other in vitro studies observed comparable biomechanical effects on the adjacent segments after fusion, but biomechanical spacer studies concentrated on the "noncompressible" X-Stop® and could not demonstrate a significant adjacent segment effect of this device.
CONCLUSIONS: The hypothesis, that an interspinous device would reduce the stress on adjacent segments compared to a rigid posterior stabilization device, could not be demonstrated with this biomechanical in vitro study. Therefore, the protection of adjacent segments after instrumentation with dynamic devices is still not completely achieved.

Entities:  

Mesh:

Year:  2011        PMID: 22094153

Source DB:  PubMed          Journal:  Acta Chir Orthop Traumatol Cech        ISSN: 0001-5415            Impact factor:   0.531


  9 in total

Review 1.  Limitations of current in vitro test protocols for investigation of instrumented adjacent segment biomechanics: critical analysis of the literature.

Authors:  David Volkheimer; Masoud Malakoutian; Thomas R Oxland; Hans-Joachim Wilke
Journal:  Eur Spine J       Date:  2015-06-03       Impact factor: 3.134

2.  Interspinous posterior devices: What is the real surgical indication?

Authors:  Alessandro Landi
Journal:  World J Clin Cases       Date:  2014-09-16       Impact factor: 1.337

3.  Preliminary efficacy of inter-spinal distraction fusion which is a new technique for lumbar disc herniation.

Authors:  Hongyu Wei; Hai Tang; Tidong Zhang; Hao Chen; Chunke Dong
Journal:  Int Orthop       Date:  2018-10-23       Impact factor: 3.075

4.  The current testing protocols for biomechanical evaluation of lumbar spinal implants in laboratory setting: a review of the literature.

Authors:  Sabrina A Gonzalez-Blohm; James J Doulgeris; William E Lee; Thomas M Shea; Kamran Aghayev; Frank D Vrionis
Journal:  Biomed Res Int       Date:  2015-02-15       Impact factor: 3.411

5.  Finite element analysis of the biomechanical effect of coflex™ on the lumbar spine.

Authors:  Dong-Hak Byun; Dong Ah Shin; Jin-Myung Kim; So-Hee Kim; Hyoung-Ihl Kim
Journal:  Korean J Spine       Date:  2012-09-30

6.  Heterotopic ossification in vertebral interlaminar/interspinous instrumentation: report of a case.

Authors:  Giuseppe Maida; Eleonora Marcati; Silvio Sarubbo
Journal:  Case Rep Surg       Date:  2012-07-24

7.  Effectiveness of dynamic fixation Coflex treatment for degenerative lumbar spinal stenosis.

Authors:  Jun-Xi Zhang; Xi-Wei Jing; Ping Cui; Xin He; Ding-Jun Hao; Shu-Jing Li
Journal:  Exp Ther Med       Date:  2017-11-13       Impact factor: 2.447

8.  Severe Retrolisthesis at the Adjacent Segment after Lumbar Fusion Combined with Dynamic Stabilization.

Authors:  Min Chan Kim; Hui Sun Wang; Chang Il Ju; Seok Won Kim
Journal:  Korean J Neurotrauma       Date:  2017-04-30

9.  Effect of lordosis on adjacent levels after lumbar interbody fusion, before and after removal of the spinal fixator: a finite element analysis.

Authors:  Fon-Yih Tsuang; Jui-Chang Tsai; Dar-Ming Lai
Journal:  BMC Musculoskelet Disord       Date:  2019-10-25       Impact factor: 2.362

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.