[Stability of volar fixed-angle plating for distal radius fractures. Failure modes in osteoporotic bone].

S Mair1, P Weninger, F Högel, S Panzer, P Augat.   

Abstract

BACKGROUND: Distal radius fractures continue to show significant complication rates after operative treatment with locked plating. Failure occurs by screw loosening or screw penetration in the distal fragment. Placement of additional screws may enhance the stiffness of fracture fixation. The aim of this study was to determine the fatigue properties of different screw configurations in distal radius plate osteosynthesis with biomechanical tests and finite element analysis (FEA).
MATERIAL AND METHODS: Unstable distal radius fractures were created in 12 human cadaveric bone specimens and were fixed with volar locking plates. Group 4SC was fixed with four screws in the distal row and group 6SC with two additional screws the row below. Dynamic loading was applied physiologically. The radial shortening, the angulation of the distal fragment and the failure mechanism were determined by experimental tests and were further elucidated by FEA.
RESULTS: Group 6SC showed a significantly lower radial shortening and inclination. Breakage of the screws within the plate was noted in group 4SC, while moderate screw penetration was observed in group 6SC. FEA confirmed the biomechanical tests. In group 4SC elevated von Mises strain in the locking mechanism explained the inclination of the screws and the distal fragment. The elastic strain in group 6SC was increased at the screw-bone interface which explained the resulting screw penetration.
CONCLUSION: The failure mechanism in volar plating of distal radius fractures depended on the number of screws and their configuration. Using two more screws increases construct stiffness and angular stability under dynamic loading. However, increased stiffness also promoted screw penetration mainly in osteoporotic bone. Compared to screw penetration, loss of reposition and inclination of the distal fragment observed in the 4SC configuration is more likely to result in clinical complications.

Entities:  

Mesh:

Year:  2013        PMID: 22072058     DOI: 10.1007/s00113-011-2099-2

Source DB:  PubMed          Journal:  Unfallchirurg        ISSN: 0177-5537            Impact factor:   1.000


  23 in total

1.  Distal radial metaphyseal forces in an extrinsic grip model: implications for postfracture rehabilitation.

Authors:  M D Putnam; N J Meyer; E W Nelson; D Gesensway; J L Lewis
Journal:  J Hand Surg Am       Date:  2000-05       Impact factor: 2.230

Review 2.  Volar plate fixation of distal radius fractures.

Authors:  Jorge Orbay
Journal:  Hand Clin       Date:  2005-08       Impact factor: 1.907

3.  Plating for distal radius fractures.

Authors:  Paul A Martineau; Gregory K Berry; Edward J Harvey
Journal:  Hand Clin       Date:  2010-02       Impact factor: 1.907

4.  Locked volar plating for unstable distal radial fractures: clinical and radiological outcomes.

Authors:  Dominique Knight; Carol Hajducka; Elizabeth Will; Margaret McQueen
Journal:  Injury       Date:  2009-10-12       Impact factor: 2.586

5.  Biomechanics of the distal radioulnar joint.

Authors:  A K Palmer; F W Werner
Journal:  Clin Orthop Relat Res       Date:  1984 Jul-Aug       Impact factor: 4.176

6.  Biomechanical evaluation of fixation of intra-articular fractures of the distal part of the radius in cadavera: Kirschner wires compared with calcium-phosphate bone cement.

Authors:  D N Yetkinler; A L Ladd; R D Poser; B R Constantz; D Carter
Journal:  J Bone Joint Surg Am       Date:  1999-03       Impact factor: 5.284

7.  Number and locations of screw fixation for volar fixed-angle plating of distal radius fractures: biomechanical study.

Authors:  Isabella Mehling; Lars P Müller; Katharina Delinsky; Dorothea Mehler; Klaus J Burkhart; Pol M Rommens
Journal:  J Hand Surg Am       Date:  2010-06       Impact factor: 2.230

Review 8.  Biomechanics and biology of plate fixation of distal radius fractures.

Authors:  Alan E Freeland; Kurre T Luber
Journal:  Hand Clin       Date:  2005-08       Impact factor: 1.907

9.  Surgical treatment of unstable distal radius fractures with volar locking plates.

Authors:  Kenichi Murakami; Yoshihiro Abe; Kazuhisa Takahashi
Journal:  J Orthop Sci       Date:  2007-03-30       Impact factor: 1.601

10.  Outcome of war-injured patients treated at first aid posts of the International Committee of the Red Cross.

Authors:  A J Korver
Journal:  Injury       Date:  1994-01       Impact factor: 2.586

View more
  1 in total

1.  [Applications of numerical simulation in musculoskeletal research and its impact on orthopedic surgery].

Authors:  D Kluess; C Hurschler; C Voigt; A Hölzer; M Stoffel
Journal:  Orthopade       Date:  2013-04       Impact factor: 1.087

  1 in total