Hassan Shaker1, Alsayed Alokda, Hisham Mahmoud.
Abstract
The objective of this work is to test the ablation capability and fiber degradation of the novel Twister fibers (TW), in both the large (LTW) and the standard (STW) sizes, against the standard side-firing (SF) fiber in a clinical setting during the treatment of BPH patients using the 980-nm high-power diode laser (HPDL). One hundred and twenty BPH patients treated with HPDL (Ceralase300, Biolitec AG, Jena, Germany) were randomized to receive treatment by one of the three fibers. Operative time corrected to tissue volume, laser treatment time, and laser energy were measured. Ablation rate was calculated as follows: the decrease of the prostate volume after 6 months/laser time. The fibers' resistance to degradation was defined by the laser energy needed to degrade the fiber completely. Preoperative prostate volume of 76 ± 38, 70 ± 39, and 88 ± 49 cc decreased by 49 ± 16, 51 ± 20, and 63 ± 16% for the SF, STW, and LTW fibers, respectively. This difference was highly significant when the LTW was compared to the other two fibers (p < 0.001). Prostate volume reduction post-operatively within each group as compared to the pre-operative volume was highly significant (p < 0.001). The ablation rate was highest in LTW, being 1.31 ± 0.59, 1.09 ± 0.51, and 1.54 ± 0.44 cc/min for the SF, STW, and LTW fibers, respectively. The LTW fiber resisted degradation more than the other fibers and the STW more than the SF fiber (p < 0.001). This study demonstrates the higher ablation efficiency and resistance to degradation of the LTW fiber as compared to the STW and SF fibers. The STW fiber has a similar ablation rate of the SF fiber but resists degradation better.Entities:
Mesh:
Year: 2011 PMID: 22071987 PMCID: PMC3414700 DOI: 10.1007/s10103-011-1017-8
Source DB: PubMed Journal: Lasers Med Sci ISSN: 0268-8921 Impact factor: 3.161
Fig. 1The different types of laser fibers used with the 980-nm high-power diode laser: a side-firing fiber, b standard Twister fiber, c large Twister fiber
Baseline prostatic volume
| Laser fiber | Mean | Range | |
|---|---|---|---|
| Baseline prostatic volume | Side firing | 70.25 ± 39.27 | 21.00–160.00 |
| Twister | 76.90 ± 38.31 | 14.00–160.00 | |
| Large Twister | 88.66 ± 49.76 | 40.00–220.00 |
No statistical significant difference was observed between the three groups
Laser-related data
| Laser fiber | Mean | Range | |
|---|---|---|---|
| Operative time | Side-firing | 72.89 ± 51.08 | 25.00–180.00 |
| Twister | 66.50 ± 37.89 | 20.00–180.00 | |
| Large Twister | 69.96 ± 45.16 | 24.00–180.00 | |
| Laser treatment time | Side-firing | 27.69 ± 15.59 | 10.50–67.63 |
| Twister | 32.16 ± 15.41 | 11.95–68.93 | |
| Large Twister | 34.18 ± 17.06 | 11.10–65.78 | |
| Energy (kJ) | Side-firing | 271.46 ± 177.36 | 78.00–715.00 |
| Twister | 309.21 ± 145.79 | 100.00–640.00 | |
| Large Twister | 442.37 ± 194.17 | 175.00–858.00 | |
| Energy needed to ablate cc tissue | Side-firing | 9.52 ± 6.49 | 3.70–29.00 |
| Twister | 11.76 ± 8.89 | 3.50–38.30 | |
| Large Twister | 8.18 ± 2.71 | 5.00–14.90 |
No statistical significant difference was observed between the three groups
Fig. 2Operative time corrected to tissue volume. a Significant difference between the SF and STW fibers. b High statistical significant difference between the SF and the LTW. c Significant difference between the STW and LTW fibers
Fig. 3Six-month post-operative prostate volume reduction. a, b Significant difference in volume reduction between LTW and other two fibers
Fig. 4Fiber tissue ablation rate
Fig. 5Fiber resistance to degradation. The energy specified for each fiber type represents the mean energy at which fibers degraded and ceased to function properly. This occurred in 25 SF, 12 STW, and ten LTW fibers. a High significant difference between the SF and STW fibers. b High statistical significant difference between the SF and the LTW. c High significant difference between the STW and LTW fibers
Correlation between operative time versus prostate size
| Variables | Baseline prostate size | ||
|---|---|---|---|
| Side-firing | Large Twister | Twister | |
| Time OR | 0.95b | 0.90b | 0.72b |
| Laser time | 0.87b | 0.85b | 0.70a |
aSignificant correlation
bHighly significant correlation by using Spearman correlation test
Numbers in the table are the r (rank) value