Literature DB >> 22068916

Radiofrequency heating induced by 7T head MRI: thermal assessment using discrete vasculature or Pennes' bioheat equation.

Astrid L H M W van Lier1, Alexis N T J Kotte, Bas W Raaymakers, Jan J W Lagendijk, Cornelis A T van den Berg.   

Abstract

PURPOSE: To evaluate and compare the maximum temperature (T(max) ) in the head after exposure to a 300 MHz radiofrequency (RF) field induced by a magnetic resonance imaging (MRI) coil using two thermal simulation methods: Pennes' bioheat equation (PBHE) and discrete vasculature (DIVA).
MATERIALS AND METHODS: The electromagnetic field induced in the head by a 7T birdcage coil was simulated using finite-difference time-domain (FDTD) and validated by MRI. The specific absorption rate (SAR) distributions normalized to the 10-gram maximum or the whole-head average were used for PBHE and DIVA simulations.
RESULTS: For all cases, the T(max) in PBHE was slightly higher than in DIVA. The T(max) was 37.9-38.4°C, depending on the simulation method or perfusion rate.
CONCLUSION: In some situations, RF exposure limited to SAR(max,10g) led to a T(max) higher than allowed by International Electrotechnical Commission (IEC) regulations. Therefore, it is advisable to use thermal simulations to evaluate RF safety of MRI. The simulation method used only slightly influenced the observed maximum temperature; the observed temperature with PBHE was higher in all situations. So PBHE is an appropriate method for RF safety assessment of MRI in the head. Using DIVA simulations, it was found unlikely that the body temperature increases significantly due to energy deposited by a head coil under normal circumstances.
Copyright © 2011 Wiley Periodicals, Inc.

Entities:  

Mesh:

Year:  2011        PMID: 22068916     DOI: 10.1002/jmri.22878

Source DB:  PubMed          Journal:  J Magn Reson Imaging        ISSN: 1053-1807            Impact factor:   4.813


  10 in total

1.  Experience with magnetic resonance imaging of human subjects with passive implants and tattoos at 7 T: a retrospective study.

Authors:  Yacine Noureddine; Andreas K Bitz; Mark E Ladd; Markus Thürling; Susanne C Ladd; Gregor Schaefers; Oliver Kraff
Journal:  MAGMA       Date:  2015-09-26       Impact factor: 2.310

2.  Parallel transmission RF pulse design with strict temperature constraints.

Authors:  Cem M Deniz; Giuseppe Carluccio; Christopher Collins
Journal:  NMR Biomed       Date:  2017-02-10       Impact factor: 4.044

3.  Measurement of SAR-induced temperature increase in a phantom and in vivo with comparison to numerical simulation.

Authors:  Sukhoon Oh; Yeun-Chul Ryu; Giuseppe Carluccio; Christopher T Sica; Christopher M Collins
Journal:  Magn Reson Med       Date:  2013-06-26       Impact factor: 4.668

4.  Predicting long-term temperature increase for time-dependent SAR levels with a single short-term temperature response.

Authors:  Giuseppe Carluccio; Mary Bruno; Christopher M Collins
Journal:  Magn Reson Med       Date:  2015-06-22       Impact factor: 4.668

5.  Short-term side-effects of brain MR examination at 7 T: a single-centre experience.

Authors:  M Cosottini; D Frosini; L Biagi; I Pesaresi; M Costagli; G Tiberi; M Symms; M Tosetti
Journal:  Eur Radiol       Date:  2014-05-10       Impact factor: 5.315

Review 6.  Magnetic resonance safety.

Authors:  Steffen Sammet
Journal:  Abdom Radiol (NY)       Date:  2016-03

7.  From complex B(1) mapping to local SAR estimation for human brain MR imaging using multi-channel transceiver coil at 7T.

Authors:  Xiaotong Zhang; Sebastian Schmitter; Pierre-Francois Van de Moortele; Jiaen Liu; Bin He
Journal:  IEEE Trans Med Imaging       Date:  2013-03-11       Impact factor: 10.048

Review 8.  Thermal modelling using discrete vasculature for thermal therapy: A review.

Authors:  H Petra Kok; Johanna Gellermann; Cornelis A T van den Berg; Paul R Stauffer; Jeffrey W Hand; Johannes Crezee
Journal:  Int J Hyperthermia       Date:  2013-06       Impact factor: 3.914

9.  Computational and experimental evaluation of the Tic-Tac-Toe RF coil for 7 Tesla MRI.

Authors:  Narayanan Krishnamurthy; Tales Santini; Sossena Wood; Junghwan Kim; Tiejun Zhao; Howard J Aizenstein; Tamer S Ibrahim
Journal:  PLoS One       Date:  2019-01-10       Impact factor: 3.240

10.  Evaluation of specific absorption rate and heating in children exposed to a 7T MRI head coil.

Authors:  Shaihan J Malik; Jeffrey W Hand; David W Carmichael; Joseph V Hajnal
Journal:  Magn Reson Med       Date:  2022-06-06       Impact factor: 3.737