Literature DB >> 22058182

Complex shaped detachable platinum coil system for the treatment of cerebral aneurysms: the Codman Trufill DCS and Trufill DCS Orbit Detachable Coil System COMPLEX Registry final results.

Bernard R Bendok1, Rudy J Rahme.   

Abstract

OBJECTIVE: To report on an electronic database with longitudinal information on the use of bare platinum coils for the endovascular occlusion of cerebral aneurysms. We report here the final analysis of the data.
METHODS: From 1 May 2004 to 10 July 2007, a global registry was conducted at 36 centers in the USA, Europe and Latin America. 303 patients were enrolled, of whom 299 patients were treated with bare detachable platinum coils (mostly Trufill Detachable Coil System (DCS) Orbit coils). Short and long term anatomical and clinical data were collected. Logistic regression analysis was used to examine the association of selected patient characteristics to outcomes of recanalization and retreatment.
RESULTS: 313 aneurysms, ruptured and unruptured, were treated in the 299 patients. Follow-up data were available on 185 patients: 150 (81%) had a 3-6 month follow-up, 63 (34%) had a 1 year follow-up and five (3%) had a 2 year follow-up. This paper reviews the status of 185 patients with follow-up data available at this time. Immediately postprocedure, 114 (58%) aneurysms were completely obliterated, 40 (20%) residual neck, 23 (12%) residual aneurysm and 19 (10%) dog ear. Of 174 aneurysms in 167 patients with angiographic follow-up, 137 (79%) remained stable or improved, nine (5%) showed aneurysm regrowth, 26 (15%) showed compaction and two (1%) had migrated. At 3-6 months, 11 of the 174 (6%) aneurysms were re-treated.
CONCLUSION: These results confirm previous outcome data of endovascular occlusion of cerebral aneurysms with Trufill DCS Orbit complex shaped detachable platinum coils. Complications related to the device were rare.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 22058182     DOI: 10.1136/neurintsurg-2011-010118

Source DB:  PubMed          Journal:  J Neurointerv Surg        ISSN: 1759-8478            Impact factor:   5.836


  2 in total

1.  One-year efficacy and safety of the Trufill DCS Orbit and Orbit Galaxy detachable coils in the endovascular treatment of intracranial aneurysms: Results from the TRULINE study.

Authors:  Benjamin Gory; Laure Huot; Roberto Riva; Paul E Labeyrie; Olivier Levrier; Ariel Lebedinsky; Hervé Brunel; Jean-Yves Gauvrit; Raphael Blanc; Emmanuel Chabert; Laurent Derex; Evelyne Emery; Arnaud Nicolas; Hubert Desal; Georges Rodesch; Francis Turjman
Journal:  Interv Neuroradiol       Date:  2017-07-04       Impact factor: 1.610

2.  The Hydrogel Endovascular Aneurysm Treatment Trial (HEAT): A Randomized Controlled Trial of the Second-Generation Hydrogel Coil.

Authors:  Bernard R Bendok; Karl R Abi-Aad; Jennifer D Ward; Jason F Kniss; Mary J Kwasny; Rudy J Rahme; Salah G Aoun; Tarek Y El Ahmadieh; Najib E El Tecle; Samer G Zammar; Rami James N Aoun; Devi P Patra; Sameer A Ansari; Jean Raymond; Henry H Woo; David Fiorella; Guilherme Dabus; Genevieve Milot; Josser E Delgado Almandoz; John A Scott; Andrew J DeNardo; Shervin R Dashti
Journal:  Neurosurgery       Date:  2020-05-01       Impact factor: 4.654