Literature DB >> 22056692

Implementation of a multi-parameter Point-of-Care-blood test analyzer reduces central laboratory testing and need for blood transfusions in very low birth weight infants.

Ludo Mahieu1, Annick Marien, Jozef De Dooy, Margo Mahieu, Hanne Mahieu, Viviane Van Hoof.   

Abstract

Blood sampling for laboratory testing is a major cause of iatrogenic blood loss and anemia in neonatal intensive care unit [NICU] patients. The objective of the study was to assess whether the implementation of a multi-parameter Point of Care Test [POCT] (Roche, Cobas b221) analyzer affected blood loss for central laboratory testing and need for red blood cell transfusion in our NICU. This was a retrospective observational cohort study in a NICU with comparison of two serial cohorts of 2 years each. Implementation of a multi-parameter POCT decreased central laboratory performed testing for bilirubin (-32% per patient) and electrolytes (-36% per patient). On average, the net blood volume taken per admitted patient for electrolyte testing decreased with 23.7% and 22.2% for bilirubin testing in the second cohort. After implementation of POCT, fewer very low birth weight infants [VLBWI] required blood transfusion (38.9% vs. 50%, p<.05) as the number of transfusion/infants decreased by 48% (1.57 vs. 2.53, p<0.01). The implementation of POCT was cost-efficient for the Belgian national health insurance, cost reduction -8.3% per neonate. We conclude that implementation of a bedside multi-parameter POCT analyzer decreases transfusions among VLBWI by reducing iatrogenic blood loss for central laboratory testing.
Copyright © 2011. Published by Elsevier B.V.

Entities:  

Mesh:

Year:  2011        PMID: 22056692     DOI: 10.1016/j.cca.2011.10.027

Source DB:  PubMed          Journal:  Clin Chim Acta        ISSN: 0009-8981            Impact factor:   3.786


  12 in total

Review 1.  Economic Evidence and Point-of-Care Testing.

Authors:  Andrew St John; Christopher P Price
Journal:  Clin Biochem Rev       Date:  2013-08

Review 2.  Point-of-care testing in the diagnosis of gastrointestinal cancers: current technology and future directions.

Authors:  Jeremy R Huddy; Melody Z Ni; Sheraz R Markar; George B Hanna
Journal:  World J Gastroenterol       Date:  2015-04-14       Impact factor: 5.742

3.  Minimizing blood loss and the need for transfusions in very premature infants.

Authors:  Brigitte Lemyre; Megan Sample; Thierry Lacaze-Masmonteil
Journal:  Paediatr Child Health       Date:  2015 Nov-Dec       Impact factor: 2.253

4.  A research protocol for developing a Point-Of-Care Key Evidence Tool 'POCKET': a checklist for multidimensional evidence reporting on point-of-care in vitro diagnostics.

Authors:  Jeremy R Huddy; Melody Ni; Stella Mavroveli; James Barlow; Doris-Ann Williams; George B Hanna
Journal:  BMJ Open       Date:  2015-07-10       Impact factor: 2.692

Review 5.  The state of point-of-care testing: a European perspective.

Authors:  Anders Larsson; Roman Greig-Pylypczuk; Albert Huisman
Journal:  Ups J Med Sci       Date:  2015-01-26       Impact factor: 2.384

6.  Bedside ABG, electrolytes, lactate and procalcitonin in emergency pediatrics.

Authors:  Prerna Batra; Ajeet Kumar Dwivedi; Neha Thakur
Journal:  Int J Crit Illn Inj Sci       Date:  2014-07

7.  Predictive factors and practice trends in red blood cell transfusions for very-low-birth-weight infants.

Authors:  Osayame A Ekhaguere; Frank H Morriss; Edward F Bell; Nadkarni Prakash; John A Widness
Journal:  Pediatr Res       Date:  2016-01-12       Impact factor: 3.756

8.  Interventions to prevent iatrogenic anemia: a Laboratory Medicine Best Practices systematic review.

Authors:  Nedra S Whitehead; Laurina O Williams; Sreelatha Meleth; Sara M Kennedy; Nneka Ubaka-Blackmoore; Sharon M Geaghan; James H Nichols; Patrick Carroll; Michael T McEvoy; Julie Gayken; Dennis J Ernst; Christine Litwin; Paul Epner; Jennifer Taylor; Mark L Graber
Journal:  Crit Care       Date:  2019-08-09       Impact factor: 9.097

Review 9.  CMOS cell sensors for point-of-care diagnostics.

Authors:  Yekbun Adiguzel; Haluk Kulah
Journal:  Sensors (Basel)       Date:  2012-07-25       Impact factor: 3.576

10.  Factors associated with red blood cell transfusions in very-low-birth-weight preterm infants in Brazilian neonatal units.

Authors:  Amelia Miyashiro Nunes dos Santos; Ruth Guinsburg; Maria Fernanda Branco de Almeida; Renato Soibelman Procianoy; Sergio Tadeu Martins Marba; Walusa Assad Gonçalves Ferri; Ligia MariaSuppo de Souza Rugolo; José Maria Andrade Lopes; Maria Elisabeth Lopes Moreira; Jorge Hecker Luz; Maria Rafaela Conde González; Jucille do Amaral Meneses; Regina Vieira Cavalcante da Silva; Vânia Olivetti Steffen Abdallah; José Luiz Muniz Bandeira Duarte; Patricia Franco Marques; Maria Albertina Santiago Rego; Navantino Alves Filho; Vera Lúcia Jornada Krebs
Journal:  BMC Pediatr       Date:  2015-09-04       Impact factor: 2.125