Literature DB >> 22051589

Penetration and spread of interferential current in cutaneous, subcutaneous and muscle tissues.

Abulkhair Beatti1, Anton Rayner, Lucy Chipchase, Tina Souvlis.   

Abstract

OBJECTIVES: To investigate the penetration depth of interferential current (IFC) through soft tissue and the area over which it spreads during clinical application.
DESIGN: A laboratory-based study of healthy participants.
SETTING: A university research laboratory. PARTICIPANTS: Twelve healthy subjects.
INTERVENTIONS: Premodulated IFC at 90 Hz and 'true' IFC at frequencies of 4, 40 and 90 Hz were applied via four electrodes, in a quadrant setting, to the distal medial thigh of each participant on separate occasions. MAIN OUTCOME MEASURE: Voltage induced by tested currents was measured at three locations (middle of the four electrodes, in line with one circuit and outside the four electrodes) and three depths (skin, subcutaneous and muscle tissues) using three Teflon-coated needle electrodes connected to a Cambridge Electronic Design data acquisition system.
RESULTS: All voltages were greater at all depths and locations compared with baseline (P<0.001): premodulated IFC [mean difference 0.112, 95% confidence interval (CI) 0.065 to 0.160], 4 Hz (mean difference 0.168, 95% CI 0.106 to 0.229), 40 Hz (mean difference 0.165, 95% CI 0.107 to 0.223) and 90 Hz (mean difference 0.162, 95% CI 0.102 to 0.221). Voltages decreased with depth. Lower voltages of all currents were recorded in the middle of the four electrodes, with the highest voltage for 'true' IFC being recorded outside the four electrodes (mean difference 0.04, 95% CI 0.01 to 0.029; P=0.011). The premodulated IFC had the highest voltage in line with one circuit.
CONCLUSIONS: IFC passed through soft tissues, with the highest voltages recorded in superficial tissue and the lowest voltages recorded in muscle. For 'true' IFC, the current spread outside the electrodes at higher voltages compared with the intersection of the four electrodes. The premodulated IFC had the highest voltage in line with one circuit. In terms of higher recorded voltages, 'true' IFCs were more efficient than the premodulated IFC when targeting deeper tissues. However, further studies with larger samples are required to confirm the results of this study. Crown Copyright Â
© 2011. Published by Elsevier Ltd. All rights reserved.

Entities:  

Mesh:

Year:  2011        PMID: 22051589     DOI: 10.1016/j.physio.2011.01.008

Source DB:  PubMed          Journal:  Physiotherapy        ISSN: 0031-9406            Impact factor:   3.358


  11 in total

1.  Interferential electrical stimulation for improved bladder management following spinal cord injury.

Authors:  Cristina Daia; Ana Maria Bumbea; Cristinel Dumitru Badiu; Camelia Ciobotaru; Gelu Onose
Journal:  Biomed Rep       Date:  2019-07-04

2.  A computational study to evaluate the activation pattern of nerve fibers in response to interferential currents stimulation.

Authors:  Mahsa Agharezaee; Amin Mahnam
Journal:  Med Biol Eng Comput       Date:  2015-04-03       Impact factor: 2.602

3.  Transcutaneous electrical stimulation on the anterior neck region: The impact of pulse duration and frequency on maximum amplitude tolerance and perceived discomfort.

Authors:  Ali Barikroo; Giselle Carnaby; Donald Bolser; Ronald Rozensky; Michael Crary
Journal:  J Oral Rehabil       Date:  2018-03-30       Impact factor: 3.837

Review 4.  Analgesia for Sheep in Commercial Production: Where to Next?

Authors:  Alison Small; Andrew David Fisher; Caroline Lee; Ian Colditz
Journal:  Animals (Basel)       Date:  2021-04-14       Impact factor: 2.752

5.  Nerve excitation using an amplitude-modulated signal with kilohertz-frequency carrier and non-zero offset.

Authors:  Leonel E Medina; Warren M Grill
Journal:  J Neuroeng Rehabil       Date:  2016-07-12       Impact factor: 4.262

Review 6.  Neuromodulation via Interferential Electrical Stimulation as a Novel Therapy in Gastrointestinal Motility Disorders.

Authors:  Judith S Moore; Peter R Gibson; Rebecca E Burgell
Journal:  J Neurogastroenterol Motil       Date:  2018-01-30       Impact factor: 4.924

7.  Technical development of transcutaneous electrical nerve inhibition using medium-frequency alternating current.

Authors:  Yushin Kim; Hang-Jun Cho; Hyung-Soon Park
Journal:  J Neuroeng Rehabil       Date:  2018-08-20       Impact factor: 4.262

8.  Effect of high-frequency alternating current transcutaneous stimulation over muscle strength: a controlled pilot study.

Authors:  Diego Serrano-Muñoz; Juan Avendaño-Coy; Cristina Simón-Martínez; Julian Taylor; Julio Gómez-Soriano
Journal:  J Neuroeng Rehabil       Date:  2018-11-12       Impact factor: 4.262

9.  Multiscale Computational Model Reveals Nerve Response in a Mouse Model for Temporal Interference Brain Stimulation.

Authors:  Jose Gomez-Tames; Akihiro Asai; Akimasa Hirata
Journal:  Front Neurosci       Date:  2021-06-30       Impact factor: 4.677

Review 10.  Analgesic Effects of Interferential Current Therapy: A Narrative Review.

Authors:  Érika Patrícia Rampazo; Richard Eloin Liebano
Journal:  Medicina (Kaunas)       Date:  2022-01-17       Impact factor: 2.430

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.