Literature DB >> 22051198

The importance of pulsed lavage on interface temperature and ligament tension force in cemented unicompartmental knee arthroplasty.

M Clarius1, J B Seeger, S Jaeger, G Mohr, R G Bitsch.   

Abstract

BACKGROUND: Mechanical loosening is the most common cause of revision in unicompartmental knee arthroplasty. We determined the effect of bone lavage on tibial cement penetration and interface temperature with controlled ligament tension forces. We presumed pulsed lavage would allow increased cement penetration compared with syringe lavage.
METHODS: Cemented unicompartmental knee arthroplasty was performed in 12 pairs of fresh-frozen knees. Lavage was performed using pulsed lavage on one side (A) and syringe lavage on the other (B). Cement penetration pressure, interface temperature, and ligament tension forces were continuously monitored during the operation. Screened radiographs were taken and cement penetration under the tibial plateau was measured.
FINDINGS: The pulsed lavage group showed a mean cement penetration area of 187.24 (SD 36.37) mm², whereas 144.29 (SD 35.74) mm(2) was measured in the group with syringe lavage. Cement penetration pressure was 13.29 (SD 8.69) kPa in Group A and 20.21 (SD 7.78) kPa in Group B. Maximum interface temperatures of 46.99°C were observed in Group A and 45.02°C in Group B.
INTERPRETATION: Our data showed pulsed lavage cleansing of the cancellous tibial bone substantially improved cement penetration compared with syringe lavage without reaching the temperature threshold for bone necrosis. We recommend the routine use of pulsed lavage to improve long-term fixation.
Copyright © 2011 Elsevier Ltd. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 22051198     DOI: 10.1016/j.clinbiomech.2011.10.006

Source DB:  PubMed          Journal:  Clin Biomech (Bristol, Avon)        ISSN: 0268-0033            Impact factor:   2.063


  4 in total

Review 1.  [Revision after unicompartmental knee arthroplasty].

Authors:  G Mohr; J Martin; M Clarius
Journal:  Orthopade       Date:  2014-10       Impact factor: 1.087

2.  Pulsatile Lavage Systems with High Impact Pressure and High Flow Produce Cleaner Cancellous Bone Prior to Cementation in Cemented Arthroplasty.

Authors:  Kevin Knappe; Rudi G Bitsch; Mareike Schonhoff; Tilman Walker; Tobias Renkawitz; Sebastian Jaeger
Journal:  J Clin Med       Date:  2021-12-24       Impact factor: 4.241

3.  Does tibial design modification improve implant stability for total knee arthroplasty? An experimental cadaver study.

Authors:  Sebastian Jaeger; Marvin Eissler; Martin Schwarze; Mareike Schonhoff; J Philippe Kretzer; Rudi G Bitsch
Journal:  Bone Joint Res       Date:  2022-04       Impact factor: 4.410

4.  Tibial cleaning method for cemented total knee arthroplasty: An experimental study.

Authors:  Peter Helwig; Lukas Konstantinidis; Anja Hirschmüller; Verena Miltenberger; Kerstin Kuminack; Norbert P Südkamp; Oliver Hauschild
Journal:  Indian J Orthop       Date:  2013-01       Impact factor: 1.251

  4 in total