Influence of in vitro hemolysis on hematological testing on Advia 2120.

G Lippi1, R Musa, P Avanzini, R Aloe, S Pipitone, F Sandei.   

Abstract

INTRODUCTION: Although there is broad knowledge on the effect of several preanalytical errors on laboratory hematology, there is no information on the reliability of routine hematological testing on hemolyzed specimens.
METHODS: K(2) EDTA-anticoagulated blood collected from 13 healthy volunteers was divided in three aliquots. Aliquot A did not undergo further manipulation, while the second and third aliquots (B and C, respectively) were passed 5 and 10 times through a small-gauge needle to produce scalar amounts of hemolysis. Hematological testing was first performed on Advia 2120. The plasma was then separated by centrifugation and tested for lactate dehydrogenase and hemolysis index (HI).
RESULTS: As compared with the nonhemolyzed aliquot, analytically and clinically significant variations were recorded in both mildly and frankly hemolyzed specimens for lactate dehydrogenase, HI, red blood cells count, hematocrit, mean corpuscular volume, mean hemoglobin content, lymphocytes, basophils, and large unstained cells. Both the platelets count and the mean platelet volume were also dramatically affected. The other parameters tested (hemoglobin, red blood cell distribution width, neutrophils, and eosinophils) did not vary significantly.
CONCLUSION: The results of routine hematological testing on mildly to frankly hemolyzed specimens might be unreliable.
© 2011 Blackwell Publishing Ltd.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 22051137     DOI: 10.1111/j.1751-553X.2011.01378.x

Source DB:  PubMed          Journal:  Int J Lab Hematol        ISSN: 1751-5521            Impact factor:   2.877


  4 in total

1.  Interference of in vitro hemolysis complete blood count.

Authors:  Gabriela de Jonge; Talita L Dos Santos; Bruno R Cruz; Mackelly Simionatto; Jeanine I M Bittencourt; Everson A Krum; Mariane F Moss; Danielle Cristyane K Borato
Journal:  J Clin Lab Anal       Date:  2018-02-03       Impact factor: 2.352

2.  Determination of hemolysis index thresholds for biochemical tests on Siemens Advia 2400 chemistry analyzer.

Authors:  Zhenhua Du; JiQin Liu; Hua Zhang; BuHe Bao; RuiQi Zhao; Ying Jin
Journal:  J Clin Lab Anal       Date:  2019-02-19       Impact factor: 2.352

3.  Effects of specimen haemolysis on complete blood count results by Abbott Alinity hq System.

Authors:  Müjgan Ercan; Emiş Deniz Akbulut; Nihayet Bayraktar; Şerif Ercan
Journal:  Biochem Med (Zagreb)       Date:  2021-10-15       Impact factor: 2.313

Review 4.  Unreliable Automated Complete Blood Count Results: Causes, Recognition, and Resolution.

Authors:  Gene Gulati; Guldeep Uppal; Jerald Gong
Journal:  Ann Lab Med       Date:  2022-09-01       Impact factor: 4.941

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.