| Literature DB >> 22048158 |
Markus Beck1, Silvia Brockhuis, Niels van der Velde, Christian Breukers, Jan Greve, Leon W M M Terstappen.
Abstract
We present a simple system for CD4 and CD8 counting for point-of-care HIV staging in low-resource settings. Automatic sample preparation is achieved through a dried reagent coating inside a thin (26 μm) counting chamber, allowing the delayed release of fluorochrome conjugated monoclonal antibodies after the filling of the chamber with whole blood by capillary flow. A custom-built image cytometer is used to capture fluorescence images representing more than 1 μl of blood. The thin layer of blood in combination with the large image area allows the use of whole blood from a finger prick without the need for dilution, lysis or cell enrichment. Automatic cell counting of CD4(+) and CD8(+) T-lymphocytes correlates well with results obtained by flow cytometry.Entities:
Mesh:
Substances:
Year: 2011 PMID: 22048158 DOI: 10.1039/c1lc20565j
Source DB: PubMed Journal: Lab Chip ISSN: 1473-0189 Impact factor: 6.799