Accuracy of a feedback device for cardiopulmonary resuscitation on a dental chair.

Nicolas Segal1, Florian Laurent, Louis Maman, Patrick Plaisance, Pascal Augustin.   

Abstract

BACKGROUND: Conflicting studies exist about the effectiveness of cardiopulmonary resuscitation (CPR) on a dental chair. In some situations, dental surgeons are obliged to perform CPR with the patient on the chair. Feedback devices are supposed to guide the compression depth in order to improve the quality of CPR, but some devices are based on an accelerometer that can theoretically report erroneous results because of the lack of rigidity of a dental chair.
OBJECTIVE: The aim of this study was to evaluate the accuracy of these devices to guide chest compressions on a dental chair.
METHODS: A prospective, randomised, crossover, equivalence/non-inferiority study was conducted to compare the values of compression depths provided by the feedback device (Real CPR Help(®), delivered by Zoll© Medical Corporation, Chelmsford, MA, USA) with the real measurements provided by the manikin (Resusci Anne(®) Advanced SkillTrainer, Laerdal Medical AS©, Norway). Chest-compression-only CPR was performed by 15 Basic Life Support instructors who carried out two rounds of continuous CPR for 2 min each. Data were analysed with a correlation test, a Bland-Altman method and a Wilcoxon test. Statistical significance was defined as p<0.05.
RESULTS: A significant difference was found between the mean depths of compression measured by the feedback device and the manikin on a dental chair and on the floor (p<0.0001). The feedback device overestimated the depth of chest compressions, and Bland-Altman analysis demonstrated poor agreement.
CONCLUSION: This study indicates that feedback devices with accelerometer technology are not sufficiently reliable to ensure adequate chest compression on dental chairs.

Entities:  

Mesh:

Year:  2011        PMID: 22045607     DOI: 10.1136/emermed-2011-200003

Source DB:  PubMed          Journal:  Emerg Med J        ISSN: 1472-0205            Impact factor:   2.740


  3 in total

1.  Efficacy of Chest Compressions Performed on Patients in Dental Chairs Versus on the Floor.

Authors:  Yoshitaka Shimizu; Takuma Sadamori; Noboru Saeki; Akari Mukai; Mitsuru Doi; Kana Oue; Mitsuhiro Yoshida; Masahiro Irifune
Journal:  Anesth Prog       Date:  2021-06-01

Review 2.  Mapping the use of simulation in prehospital care - a literature review.

Authors:  Anna Abelsson; Ingrid Rystedt; Björn-Ove Suserud; Lillemor Lindwall
Journal:  Scand J Trauma Resusc Emerg Med       Date:  2014-03-28       Impact factor: 2.953

3.  Effect of the Cardio First Angel™ device on CPR indices: a randomized controlled clinical trial.

Authors:  Amir Vahedian-Azimi; Mohammadreza Hajiesmaeili; Ali Amirsavadkouhi; Hamidreza Jamaati; Morteza Izadi; Seyed J Madani; Seyed M R Hashemian; Andrew C Miller
Journal:  Crit Care       Date:  2016-05-17       Impact factor: 9.097

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.