Gerhard Fusch1, Arum Choi, Niels Rochow, Christoph Fusch.
Abstract
A sensitive, accurate, and specific quantitative UPLC-MS/MS method was developed for lactose measurement of cow's and human milk and validated with cow's milk samples certified by an external laboratory. The new method employs only a dilution of raw cow's and human milk for simple preparation with no need to remove protein and fat prior to analysis with UPLC-MS/MS. It was operated in negative mode to detect lactose molecules and labeled (13)C(12)-lactose with the highest sensitivity. The principle advantages of the new LC-MS/MS method were: completed lactose determination in 5 min, absolute recovery of 97-107%, lower limit of detection <5 ng/L, and 99% linearity over the concentration range of 0.7-4.4 mg/L for both cow's and human milk. The mean lactose concentration of 51 human milk samples was measured as 56.8 ± 5.5 g/L ranging from 43 to 65 g/L. The described method represents validated lactose analysis with high accuracy and precision for a routine lactose determination in raw human milk. CrownEntities:
Mesh:
Substances:
Year: 2011 PMID: 22041090 DOI: 10.1016/j.jchromb.2011.09.053
Source DB: PubMed Journal: J Chromatogr B Analyt Technol Biomed Life Sci ISSN: 1570-0232 Impact factor: 3.205