AIM: To evaluate the effects of repeated autoclaving on torsional strength of two nickel-titanium (NiTi) rotary endodontic files: Twisted Files (SybronEndo, Orange, CA, USA) and GT Series X files (Dentsply Tulsa Dental Specialties, Tulsa, OK, USA).
METHODOLOGY: Four groups of 20 GT Series X (size 20, .06 taper) and four groups of 20 Twisted Files (size 25, .06 taper) were subjected to 0, 1, 3 or 7 autoclave cycles. The nonautoclaved files served as controls. Failure in torsion was recorded using a torsiometer according to ANSI/ADA 58 and ISO 3630-1 standards. Mean results were analysed using analysis of variance (anova) at a 95% confidence level. Dunnett's procedure was used for post hoc analysis.
RESULTS: There was no significant difference in torsional moment between the number of autoclave cycles for Twisted Files. However, the mean torque at failure was significantly lower for GT Series X files after three (P<0.001) and seven (P<0.001) autoclave cycles. For Twisted Files, there was a significant increase (P<0.001) in mean number of degrees of rotation to failure with more autoclave cycles compared to nonautoclaved files. For GT Series X files, there was no significant association (P=0.527) between the number of autoclave cycles and the degrees of rotation to failure.
CONCLUSIONS: Repeat autoclaving of unused GT Series X files between three and seven times resulted in a significant reduction in torsional strength, whilst there was no effect observed for Twisted Files.
© 2011 International Endodontic Journal.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 22034927     DOI: 10.1111/j.1365-2591.2011.01958.x

Source DB:  PubMed          Journal:  Int Endod J        ISSN: 0143-2885            Impact factor:   5.264


  5 in total

1.  Effect of autoclave sterilization on cyclic fatigue and torsional fracture resistance of NiTi rotary instruments.

Authors:  Wooyoung Kim; Soram Oh; Gil-Joo Ryu; Tae-Hwan Kim; Sung-Jae Kim; Dong-Hyung Kim; Bin-Na Lee; Kee-Yeon Kum; Seok Woo Chang; Ji-Hyun Jang
Journal:  Odontology       Date:  2019-09-11       Impact factor: 2.634

2.  The effect of multiple autoclave cycles on the surface of rotary nickel-titanium endodontic files: An in vitro atomic force microscopy investigation.

Authors:  Ashish Shashikant Nair; Mahima Tilakchand; Balaram Damodar Naik
Journal:  J Conserv Dent       Date:  2015 May-Jun

Review 3.  Nickel-Titanium Rotary File Systems: What's New?

Authors:  Sadia Tabassum; Kamil Zafar; Fahad Umer
Journal:  Eur Endod J       Date:  2019-10-18

4.  The Effect of Autoclave Sterilization on Resistance to Cyclic Fatigue of Hero Endodontic File #642 (6%) at Two Artificial Curvature.

Authors:  Masoud Khabiri; Maziar Ebrahimi; Mohammad Reza Saei
Journal:  J Dent (Shiraz)       Date:  2017-12

5.  Effect of Sterilization on Cyclic Fatigue Resistance of Proflexendo Endodontic Rotary Files.

Authors:  Emad O Alshwaimi