Literature DB >> 22034832

Clinical utility of Abbott Precision Xceed Pro® ketone meter in diabetic patients.

Hoi-Ying Elsie Yu1, Michael Agus, Mark D Kellogg.   

Abstract

BACKGROUND: Diagnosis and management of diabetic ketoacidosis (DKA) often rely on the measurement of urine ketones along with blood glucose, anion gap, and pH. These values, however, do not reliably reflect the severity of ketoacidosis. The Abbott Precision Xceed Pro® meter is an FDA-approved device that quantitatively measures β-hydroxybutyrate (BOH) in whole blood. This study was undertaken to determine whether the ketone meter meets the analytical criteria to aid DKA diagnosis and management in the hospital.
METHODS: 54 heparinized venous whole blood BOH concentrations from 27 diabetic patients were measured by the Abbott meter, and compared with the plasma BOH concentrations measured with Stanbio reagent (reference method). Measurements were done in the hospital central laboratory.
RESULTS: Of the 54 pairs of specimens analyzed, 17 pairs displayed a difference of >15% between the two methods. Nearly all discrepant points occurred when BOH >5 mmol/L (reference method). Linearity evaluation revealed that the meter is not linear from 0.0 to 8.0 mmol/L, contrary to the claim by the manufacturer. Further, we identified acetoacetate, a metabolite commonly present in DKA patients, as a potential interfering substance for the meter BOH measurement.
CONCLUSIONS: BOH measurements by the Abbott meter up to 3 mmol/L correlate well with the reference method, but become discrepant above that point. While this characteristic may be useful in the diagnosis of DKA, it may not allow clinicians to serially follow the response to therapy in hospitalized DKA patients with BOH values greater than 5 mmol/L (reference method).
© 2011 John Wiley & Sons A/S.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 22034832     DOI: 10.1111/j.1399-5448.2011.00768.x

Source DB:  PubMed          Journal:  Pediatr Diabetes        ISSN: 1399-543X            Impact factor:   4.866


  10 in total

1.  Evaluation of Nova StatStrip and FreeStyle Precision Pro blood ketone tests using 3-hydroxybutyrate doped samples.

Authors:  Johanna Helmersson-Karlqvist; Kristina Höög Hammarström; Kicki Palmberg; Carolina Backman-Johansson
Journal:  J Clin Lab Anal       Date:  2019-02-27       Impact factor: 2.352

2.  A Paper-Based "Pop-up" Electrochemical Device for Analysis of Beta-Hydroxybutyrate.

Authors:  Chien-Chung Wang; Jonathan W Hennek; Alar Ainla; Ashok A Kumar; Wen-Jie Lan; Judy Im; Barbara S Smith; Mengxia Zhao; George M Whitesides
Journal:  Anal Chem       Date:  2016-05-31       Impact factor: 6.986

Review 3.  Blood Ketones: Measurement, Interpretation, Limitations, and Utility in the Management of Diabetic Ketoacidosis.

Authors:  Ketan Dhatariya
Journal:  Rev Diabet Stud       Date:  2017-02-10

4.  Blood versus urine ketone monitoring in a pediatric cohort of patients with type 1 diabetes: a crossover study.

Authors:  Line Goffinet; Thierry Barrea; Véronique Beauloye; Philippe A Lysy
Journal:  Ther Adv Endocrinol Metab       Date:  2016-12-13       Impact factor: 3.565

5.  Plasma β-Hydroxybutyrate for the Diagnosis of Diabetic Ketoacidosis in the Emergency Department.

Authors:  Elise Schlissel Tremblay; Kate Millington; Michael C Monuteaux; Richard G Bachur; Joseph I Wolfsdorf
Journal:  Pediatr Emerg Care       Date:  2021-12-01       Impact factor: 1.454

6.  An Intermittent Fasting Mimicking Nutrition Bar Extends Physiologic Ketosis in Time Restricted Eating: A Randomized, Controlled, Parallel-Arm Study.

Authors:  Angie W Huang; Min Wei; Sara Caputo; Melissa L Wilson; Joseph Antoun; William C Hsu
Journal:  Nutrients       Date:  2021-04-30       Impact factor: 5.717

7.  Controversies Around the Measurement of Blood Ketones to Diagnose and Manage Diabetic Ketoacidosis.

Authors:  Eric S Kilpatrick; Alexandra E Butler; Linda Ostlundh; Stephen L Atkin; David B Sacks
Journal:  Diabetes Care       Date:  2022-02-01       Impact factor: 19.112

Review 8.  Evaluation of the Accuracy of Capillary Hydroxybutyrate Measurement Compared with Other Measurements in the Diagnosis of Diabetic Ketoacidosis: A Systematic Review.

Authors:  Joanne Brooke; Marlon Stiell; Omorogieva Ojo
Journal:  Int J Environ Res Public Health       Date:  2016-08-23       Impact factor: 3.390

Review 9.  Pregnancy Ketonemia and Development of the Fetal Central Nervous System.

Authors:  Agata Bronisz; Mateusz Ozorowski; Magdalena Hagner-Derengowska
Journal:  Int J Endocrinol       Date:  2018-06-04       Impact factor: 3.257

10.  Reliability and diagnostic performance of a new blood ketone and glucose meter in humans.

Authors:  Andrew Ray Moore; Angelia Maleah Holland-Winkler; Jenna Kate Ansley; Eric Deiondre Hunter Boone; Megahn Kimberanne O'Reilly Schulte
Journal:  J Int Soc Sports Nutr       Date:  2021-01-07       Impact factor: 5.150

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.