| Literature DB >> 22034617 |
Cheng-Xiong Yang1, Shan-Shan Liu, He-Fang Wang, Shan-Wei Wang, Xiu-Ping Yan.
Abstract
Metal-organic framework MIL-53(Al) was explored as the stationary phase for high-performance liquid chromatographic separation of position isomers using a binary and/or polar mobile phase. Baseline separations of xylene, dichlorobenzene, chlorotoluene and nitrophenol isomers were achieved on the slurry-packed MIL-53(Al) column with high resolution and good precision. The effects of mobile phase composition, injected sample mass and temperature were investigated. The separation of xylene, dichlorobenzene, chlorotoluene and nitrophenol isomers on MIL-53(Al) were controlled by entropy change. This journal is © The Royal Society of Chemistry 2012Entities:
Mesh:
Substances:
Year: 2011 PMID: 22034617 DOI: 10.1039/c1an15600d