| Literature DB >> 22034561 |
David L Demill1, Majid Moshirfar, Marcus C Neuffer, Maylon Hsu, Shameema Sikder.
Abstract
BACKGROUND: To compare the average values of the American Society of Cataract and Refractive Surgery (ASCRS) and Ocular MD intraocular lens (IOL) calculators to assess their accuracy in predicting IOL power in patients with prior laser-in-situ keratomileusis (LASIK) or photorefractive keratectomy.Entities:
Keywords: American Society of Cataract and Refractive Surgery; intraocular lens calculator; laser-in-situ keratomileusis; ocular MD; photorefractive keratectomy
Year: 2011 PMID: 22034561 PMCID: PMC3198415 DOI: 10.2147/OPTH.S24131
Source DB: PubMed Journal: Clin Ophthalmol ISSN: 1177-5467
Patient demographics
| Data | Mean ± standard deviation | Range |
|---|---|---|
| Age (years) | 62.48 ± 9.55 | 42 to 82 |
| Pre-LASIK/PRK SE (D) | −5.55 ± 2.50 | −10.00 to −2.00 |
| Pre-LASIK/PRK average K (D) | 44.04 ± 1.67 | 41.20 to 47.40 |
| Post-LASIK/PRK SE (D) | −0.38 ± 0.72 | −2.00 to 0.63 |
| Post-LASIK/PRK average K (D) | 40.16 ± 2.07 | 37.00 to 44.87 |
| Pre-cataract surgery SE (D) | −2.42 ± 1.77 | −6.50 to −0.13 |
| Pre-cataract surgery average K (D) | 40.03 ± 2.34 | 36.09 to 44.87 |
| Axial Length (mm) | 26.10 ± 1.18 | 24.11 to 27.71 |
| Anterior chamber depth (mm) | 3.48 ± 0.30 | 3.02 to 4.20 |
| IOL Power (D) | 20.36 ± 2.09 | 16.50 to 24.50 |
| Post-cataract surgery SE (D) | −0.56 ± 0.77 | −2.50 to 0.63 |
Abbreviations: SE, spherical equivalent; D, diopters; K, keratometry.
Figure 1Boxplot of IOL power prediction errors with different methods.
Figure 2Modified Bland–Altman plots for each method illustrating IOL prediction errors for all 21 patients. The red line represents a mean IOL prediction error of zero, the green lines represent an IOL prediction error of ±0.71 (IOL refractive prediction error ±0.50 diopters), and the blue lines represent an IOL prediction error of ±1.43 (refractive prediction error of ±1.00 diopters). The black lines on each plot represent the mean and two standard deviations for that specific method.
Mean arithmetic and absolute IOL prediction errors and ranges, variance of mean arithmetic IOL prediction errors, and percentage of eyes within ±0.50 and ±1.00 diopters (D) for different methods
| N = 21 eyes | IOL Prediction error | Percentage within | |||||
|---|---|---|---|---|---|---|---|
| Arithmetic mean ± SD | Arithmetic range | Absolute mean ± SD | Absolute range | Variance | ±0.50 D | ±1.00 D | |
| ASCRS | 0.11 ± 0.82 | −0.93 to 2.40 | 0.61 ± 0.54 | 0.08 to 2.40 | 0.66 | 71.43 | 90.48 |
| Ocular MD SRK/T | 0.57 ± 0.74 | −1.03 to 2.29 | 0.70 ± 0.61 | 0.01 to 2.29 | 0.55 | 52.38 | 95.24 |
| Ocular MD Haigis | −0.61 ± 0.72 | −2.28 to 0.69 | 0.77 ± 0.54 | 0 to 2.28 | 0.52 | 52.38 | 90.48 |
| Ocular MD average | −0.02 ± 0.68 | −1.66 to 1.49 | 0.51 ± 0.44 | 0 to 1.66 | 0.46 | 76.19 | 90.48 |
| All calculator average | 0.02 ± 0.68 | −1.37 to 1.48 | 0.50 ± 0.45 | 0.02 to 1.48 | 0.47 | 76.19 | 95.24 |