Dong Hyun Kim1, Seong-Woong Kang, Yoon Ghil Park, Won Ah Choi, Hye Ree Lee.
Abstract
PURPOSE: For patients with neuromuscular disease, air stacking, which inflates the lungs to deep volumes, is important for many reasons. However, neuromuscular patients with severe glottic dysfunction or indwelling tracheostomy tubes cannot air stack effectively. For these patients, we developed a device that permits deep lung insufflations substituting for glottic function.Entities:
Mesh:
Year: 2011 PMID: 22028162 PMCID: PMC3220243 DOI: 10.3349/ymj.2011.52.6.972
Source DB: PubMed Journal: Yonsei Med J ISSN: 0513-5796 Impact factor: 2.759
Fig. 1A fully assembled apparatus for measuring LIC with the artificial external glottic device (LICA): a resuscitation bag is connected to the AEGD.
Connection part (A); Control part (B). LIC, lung insufflation capacity; AEGD, artificial external glottic device; LICA, LIC with AGED.
Fig. 2Connection part of the artificial external glottic device (AEGD). (A) Disconnected view of the connection part. (B) Photograph of the connection part.
Fig. 3Diagram of the control part of the artificial external glottic device (AEGD). (A) Demounted view of the control part: control part is connected to the connection part through (a), and measurement apparatus such as a spirometer through (b), the external glottic function is controlled by an externally exposed part (c) of the pushing bar a', air flows from (a) to (b) like an opening of a normal glottis. (B) Photograph of the control part of AEGD: ports connecting to the connection part (a); to the measurement apparatus such as a spirometer (b); and to the external control of the glottis (c).
Fig. 4Actual application of the apparatus for the measurement of LICA. LICA, LIC with AEGD.
Comparison of Pulmonary Function Parameters Under Diagnostic Classification
VC, vital capacity; LICA, lung insufflation capacity with artificial external glottic device; SD, standard deviation; ALS, amyotrophic lateral sclerosis; SCI, spinal cord injury.
Mean values±SD (mL).
Comparison of Pulmonary Function Test Parameters Under Different Conditions
VC, vital capacity; MIC, maximum insufflation capacity; LICA, lung insufflation capacity with the artificial external glottic device; SD, standard deviation; UC, uncheckable.
Mean values±SD (mL).
*p<0.05, comparison between VC and LICA in group 2.