The impact on drug mass flow rate of interrupting and resuming carrier fluid flow: an in vitro study on a very low dead-space volume infusion set.

Damien Lannoy1, Bertrand Decaudin, Nicolas Simon, Christine Barthelemy, Bertrand Debaene, Pascal Odou.   

Abstract

BACKGROUND: Stopping and resuming carrier fluid flow can lead to potentially dangerous transient disturbances in drug mass flow rate. We compared the impact of 2 infusion sets, one with very low dead-space volume and the other with greater dead-space volume, on the amount of drug delivered during stop-and-go carrier fluid flows.
METHODS: Two infusion sets, both with antireflux, connected to an angiocatheter and with dead-space volumes of 6.185 mL and 0.071 mL, respectively, were assessed. Two protocols were studied: carrier fluid flow of 90 mL/h associated with noradrenaline infused at 7 mL/h and carrier fluid flow of 350 mL/h with a noradrenaline infusion flow of 65 mL/h. During both protocols, the carrier fluid was stopped and resumed at the same rate 30 minutes later. Effluent noradrenaline concentration was measured using UV spectrophotometry. Flow change efficiency was calculated from the ratio of the area under the experimental mass flow rate curve to the area under the theoretical instantaneous mass flow rate curve.
RESULTS: For both flow rate conditions, flow change efficiency was significantly different for the 2 infusion sets during the 10-minute period after stopping carrier fluid flow and the 10-minute period after restarting it. The major phenomena were sudden decreases in drug delivery after stopping carrier flow and sudden, temporary increases when it was resumed. The very low dead-space volume infusion set resulted in significant reduction in changes in drug delivery compared with the standard set, even at high flow rates.
CONCLUSION: The use of a very low dead-space volume set attenuates disturbances in drug delivery caused by interrupting and resuming carrier fluid flow.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 22025488     DOI: 10.1213/ANE.0b013e3182373a27

Source DB:  PubMed          Journal:  Anesth Analg        ISSN: 0003-2999            Impact factor:   5.108


  3 in total

1.  Influence of vancomycin infusion methods on endothelial cell toxicity.

Authors:  Maryline Drouet; Feng Chai; Christine Barthélémy; Gilles Lebuffe; Bertrand Debaene; Bertrand Décaudin; Pascal Odou
Journal:  Antimicrob Agents Chemother       Date:  2014-11-24       Impact factor: 5.191

2.  Effect of insulin infusion line on glycaemic variability in a perioperative high dependency unit (HDU): a prospective randomised controlled trial.

Authors:  Stéphanie Genay; Bertrand Décaudin; Sabine Ethgen; Arnaud Alluin; Elodie Babol; Julien Labreuche; Hélène Behal; Marie-Christine Vantyghem; Pascal Odou; Gilles Lebuffe
Journal:  Ann Intensive Care       Date:  2017-07-11       Impact factor: 6.925

3.  Disturbance of Vancomycin Infusion Flow during Multidrug Infusion: Influence on Endothelial Cell Toxicity.

Authors:  Maryline Drouet; Elodie Cuvelier; Feng Chai; Stéphanie Genay; Pascal Odou; Bertrand Décaudin
Journal:  Antibiotics (Basel)       Date:  2021-12-24
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.