Literature DB >> 22024897

Application of intraoperative computed tomography with or without navigation system in surgical correction of spinal deformity: a preliminary result of 59 consecutive human cases.

Geng Cui1, Yan Wang, Ting-Hsien Kao, Yonggang Zhang, Zhengsheng Liu, Baowei Liu, Jie Li, Xifeng Zhang, Shourong Zhu, Ning Lu, Keya Mao, Zheng Wang, Xuesong Zhang, Xincheng Yuan, Tianxiang Dong, Songhua Xiao.   

Abstract

STUDY
DESIGN: A retrospective analysis of patients undergoing spinal deformity correction surgery by the assistance of intraoperative computed tomography (iCT) with or without navigation system.
OBJECTIVE: To share our preliminary experience and analysis of the iCT navigation system applied to spinal deformity surgery. SUMMARY OF BACKGROUND DATA: The iCT navigation system has been shown to improve accuracy and safety in posterior instrumentation. It not only decreased the operation time but also prevented excessive radiation exposure to the medical staff. To date, there are only few reports about the application of the iCT navigation system in spinal deformity surgery.
METHODS: From April 2009 to September 2010, 59 patients who had a diagnosis of scoliosis, kyphosis, or scoliokyphosis and underwent iCT-assisted surgical correction were included. Without randomization, 28 patients were operated with the iCT-navigation system, and the other 31 patients were operated with standard procedure under iCT assistance. The detailed procedures, preoperative and intraoperative images were illustrated. The accuracy of screw placement, time for screw insertion, postoperative correction rate, and iCT scanning data were analyzed.
RESULTS: There were significant differences between 2 groups in (1) the preoperative Cobb angle (76.2° and 62.6° in the navigation and non-navigation groups), (2) the accuracy and the revision rate of thoracic pedicle screws and total pedicle screws, and (3) the average screw insertion time. The breach rate and the revision rate of thoracic pedicle screws and total pedicle screws were significantly lower and the average screw insertion time was significantly lesser in the navigation group than in the non-navigation group. There were no statistically significant difference in (1) the breach rate and the revision rate of lumbar pedicle screws, (2) the mean iCT scanning time and time-out, (3) the mean number of fusion segments, (4) the mean number of iCT scans, and (5) the postoperative correction rate. Complications were encountered in 2 patients in the non-navigation group but none in the navigation group. There was no reoperation due to implant malposition in both groups.
CONCLUSION: The iCT navigation system provides desirable accuracy of posterior spinal instrumentation for patients during surgical correction of spinal deformity without radiation exposure to the medical staff, especially in thoracic spine instrumentation. Meanwhile, the iCT in itself is an effective means of assessing complex instrumentation of the spinal deformity.

Entities:  

Mesh:

Year:  2012        PMID: 22024897     DOI: 10.1097/BRS.0b013e31823aff81

Source DB:  PubMed          Journal:  Spine (Phila Pa 1976)        ISSN: 0362-2436            Impact factor:   3.468


  26 in total

Review 1.  Osteotomy of the spine for multifocal deformities.

Authors:  Ibrahim Obeid; Louis Boissière; Jean-Marc Vital; Anouar Bourghli
Journal:  Eur Spine J       Date:  2014-11-13       Impact factor: 3.134

2.  Interest of intra-operative 3D imaging in spine surgery: a prospective randomized study.

Authors:  Sébastien Ruatti; C Dubois; E Chipon; G Kerschbaumer; M Milaire; A Moreau-Gaudry; J Tonetti; Ph Merloz
Journal:  Eur Spine J       Date:  2015-07-26       Impact factor: 3.134

3.  Tracked ultrasound snapshots in percutaneous pedicle screw placement navigation: a feasibility study.

Authors:  Tamas Ungi; Eric Moult; Joseph H Schwab; Gabor Fichtinger
Journal:  Clin Orthop Relat Res       Date:  2013-08-17       Impact factor: 4.176

Review 4.  Techniques and accuracy of thoracolumbar pedicle screw placement.

Authors:  Varun Puvanesarajah; Jason A Liauw; Sheng-Fu Lo; Ioan A Lina; Timothy F Witham
Journal:  World J Orthop       Date:  2014-04-18

Review 5.  Methods to determine pedicle screw placement accuracy in spine surgery: a systematic review.

Authors:  Ahmed A Aoude; Maryse Fortin; Rainer Figueiredo; Peter Jarzem; Jean Ouellet; Michael H Weber
Journal:  Eur Spine J       Date:  2015-03-07       Impact factor: 3.134

6.  Accuracy of pedicle screw insertion by AIRO® intraoperative CT in complex spinal deformity assessed by a new classification based on technical complexity of screw insertion.

Authors:  S Rajasekaran; Manindra Bhushan; Siddharth Aiyer; Rishi Kanna; Ajoy Prasad Shetty
Journal:  Eur Spine J       Date:  2018-01-09       Impact factor: 3.134

Review 7.  Complex osteotomies vertebral column resection and decancellation.

Authors:  Ibrahim Obeid; Anouar Bourghli; Louis Boissière; Jean-Marc Vital; Cédric Barrey
Journal:  Eur J Orthop Surg Traumatol       Date:  2014-05-16

8.  Current concepts and controversies on adolescent idiopathic scoliosis: Part II.

Authors:  Alok Sud; Athanasios I Tsirikos
Journal:  Indian J Orthop       Date:  2013-05       Impact factor: 1.251

9.  Lumbosacral pedicle screw placement using a fluoroscopic pedicle axis view and a cannulated tapping device.

Authors:  Toshitaka Yoshii; Takashi Hirai; Tsuyoshi Yamada; Satoshi Sumiya; Renpei Mastumoto; Tsuyoshi Kato; Mitsuhiro Enomoto; Hiroyuki Inose; Shigenori Kawabata; Kenichi Shinomiya; Atsushi Okawa
Journal:  J Orthop Surg Res       Date:  2015-05-28       Impact factor: 2.359

10.  Position and complications of pedicle screw insertion with or without image-navigation techniques in the thoracolumbar spine: a meta-analysis of comparative studies.

Authors:  Jinshan Tang; Ziqiang Zhu; Tao Sui; Dechao Kong; Xiaojian Cao
Journal:  J Biomed Res       Date:  2014-04-12
View more