Literature DB >> 22015223

[Impact of digital mammography in the detection and management of microcalcifications].

M Mellado1, A M Osa, A Murillo, R Bermejo, A Burguete, M J Pons, N Erdozain.   

Abstract

OBJECTIVE: To determine whether the introduction of digital mammography in breast cancer screening has resulted in changes in the detection and management of microcalcifications.
MATERIAL AND METHODS: We retrospectively studied the performance indicators of our breast cancer screening program that are related to the diagnosis of microcalcifications (rates of recall and recommendation of intermediate follow-up after screening, rate of indication of invasive procedures for microcalcifications and their positive predictive value, detection rate for microcalcifications, and number of ductal carcinomas in situ (DCIS) diagnosed). We compared the results obtained using direct digital mammography (september 2008-august 2009) with those obtained using analog mammography (September 2006-August 2007). STATISTICAL ANALYSIS: Chi-square test and measures of association.
RESULTS: We found that using digital mammography led to significant increases in the recall rate (from 50.8 to 64‰), in the rate of intermediate follow-up after screening (from 9.41 to 18.7‰), in the rate of indication for invasive procedures (from 1.88 to 3.01‰), in the cancers detected through microcalcifications (from 0.86 to 1.36‰), and in the number of DCIS diagnosed.
CONCLUSION: Direct digital mammography has improved the detection of microcalcifications, increasing the number of DCIS diagnosed without decreasing the positive predictive value of the invasive procedures indicated for microcalcifications. However, direct digital mammography has had a negative effect by increasing the recall rate and indication for short-term follow-up, possibly due to the difficulty of comparing the findings with those of earlier analog mammograms.
Copyright © 2011 SERAM. Published by Elsevier Espana. All rights reserved.

Entities:  

Mesh:

Year:  2011        PMID: 22015223     DOI: 10.1016/j.rx.2011.06.011

Source DB:  PubMed          Journal:  Radiologia        ISSN: 0033-8338


  1 in total

Review 1.  Comparison of Diagnostic Test Accuracy of Cone-Beam Breast Computed Tomography and Digital Breast Tomosynthesis for Breast Cancer: A Systematic Review and Meta-Analysis Approach.

Authors:  Temitope Emmanuel Komolafe; Cheng Zhang; Oluwatosin Atinuke Olagbaju; Gang Yuan; Qiang Du; Ming Li; Jian Zheng; Xiaodong Yang
Journal:  Sensors (Basel)       Date:  2022-05-09       Impact factor: 3.847