Toshiki Kobayashi1, Aaron Kl Leung, Stephen W Hutchins.
Abstract
BACKGROUND AND AIM: Ankle joint stiffness and its range of motion (ROM) are commonly assessed to determine the appropriate mechanical characteristics required in an effective ankle-foot orthosis (AFO) prescription. The aim of this technical note is to present the design of a manual device that enables their convenient measurement in the clinical setting and to demonstrate its reliability. TECHNIQUE: The manual device was designed with a torquemeter, a potentiometer, a steering wheel, a rotary plate, and a foot plate. The measurement of resistive torque at 0° (neutral), 5° (dorsiflexion) and 10° (dorsiflexion) ankle angular positions demonstrated the high reliability of the device with Intraclass Correlation Coefficient (ICC) (1,k) values over 0.97. DISCUSSION: Quantitative measurement of ankle joint stiffness and ROM by this manual device would provide objective information that could potentially assist AFO prescriptions. A future study should investigate how to incorporate the measurement obtained from the device into the prescription of an AFO.Entities:
Mesh:
Year: 2011 PMID: 22013156 DOI: 10.1177/0309364611425696
Source DB: PubMed Journal: Prosthet Orthot Int ISSN: 0309-3646 Impact factor: 1.895