Literature DB >> 22011242

Point-of-care measurement of serum creatinine in the intensive care unit.

Paolo Calzavacca1, Augustine Tee, Elisa Licari, Antoine Guillaume Schneider, Rinaldo Bellomo.   

Abstract

OBJECTIVE: To test the precision and limits of agreement of point-of-care testing (POCT)-based measurement of serum creatinine (Cr) in critically ill patients.
METHODS: We studied 250 paired blood samples from 82 critically ill patients from a general intensive care unit by simultaneous POCT and central laboratory testing (Jaffé method). Correlation, precision, bias, and limits of agreement were assessed. Possible confounders for interference of noncreatinine chromogens were evaluated by multivariate linear regression analysis.
RESULTS: The mean difference in serum Cr measured by central laboratory and POCT was +9.6 μmol/L (95% limits of agreement: -11.2 to +30.4 μmol/L). The mean percentage difference between the two techniques was 8.7% (95% limits of agreement -7.8% to +25.1%). On multivariate regression, the difference in serum Cr was increased with greater hemoglobin and lactate levels but decreased with greater bilirubin, albumin, and calcium levels.
CONCLUSIONS: Compared with the central laboratory testing, POCT-based measurement of serum Cr in critically ill patients carried a small negative bias. This difference appeared affected by the blood levels of biochemical variables known to affect the Jaffé method. POCT-based Cr measurement appears sufficiently accurate for clinical use.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 22011242     DOI: 10.3109/0886022X.2011.623558

Source DB:  PubMed          Journal:  Ren Fail        ISSN: 0886-022X            Impact factor:   2.606


  4 in total

1.  The nature and discriminatory value of urinary neutrophil gelatinase-associated lipocalin in critically ill patients at risk of acute kidney injury.

Authors:  Neil J Glassford; Antoine G Schneider; Shengyuan Xu; Glenn M Eastwood; Helen Young; Leah Peck; Per Venge; Rinaldo Bellomo
Journal:  Intensive Care Med       Date:  2013-08-06       Impact factor: 17.440

2.  Acute Kidney Injury Recognition in Low- and Middle-Income Countries.

Authors:  Jorge Cerdá; Sumit Mohan; Guillermo Garcia-Garcia; Vivekanand Jha; Srinivas Samavedam; Swarnalata Gowrishankar; Arvind Bagga; Rajasekara Chakravarthi; Ravindra Mehta
Journal:  Kidney Int Rep       Date:  2017-04-25

3.  Evaluation of a point-of-care serum creatinine measurement device and the impact on diagnosis of acute kidney injury in pediatric cardiac patients: A retrospective, single center study.

Authors:  Satoshi Kimura; Tatsuo Iwasaki; Kazuyoshi Shimizu; Tomoyuki Kanazawa; Hirokazu Kawase; Naohiro Shioji; Yasutoshi Kuroe; Satoshi Isoyama; Hiroshi Morimatsu
Journal:  Health Sci Rep       Date:  2019-11-24

4.  Point of care creatinine derived eGFR measurement in capillary blood for identifying patients at risk.

Authors:  Sandrine Lemoine; Anne-Cécile Rouveure; Laurence Dubourg; Solenne Pelletier; Christelle Marolho; Evelyne Decullier; Maurice Laville
Journal:  Pract Lab Med       Date:  2022-07-14
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.