Literature DB >> 22010972

Five-year follow-up of intracardiac echocardiography-assisted transcatheter closure of complex ostium secundum atrial septal defect.

Gianluca Rigatelli1, Fabio Dell' Avvocata, Paolo Cardaioli, Massimo Giordan, Dobrin Vassiliev, Nguyen T Nghia, Jack P Chen.   

Abstract

OBJECTIVE: We sought to prospectively evaluate long-term follow-up results of intracardiac echocardiography-aided transcatheter closure of complex atrial septal defects (ASD) in the adults. DESIGN AND SETTINGS: Prospective multicenter registry in tertiary care hospitals. PATIENTS AND
INTERVENTIONS: Over a 5-year period, we prospectively enrolled 56 patients (mean age 49 ± 16.7 years, 24 females) who have been referred to our center for catheter-based closure of complex secundum ASD (> 25 mm diameter, deficiency of ≥ 1 rim, multiple secundum ASD, multiperforated ASD, associated incomplete floor of the fossa ovalis with or without aneurysm, embryonic remnants of incomplete atrial septation). All patients were screened by means of transesophageal echocardiography before the operation. Eligible patients underwent intracardiac echocardiography study and closure attempt.
RESULTS: Forty patients underwent a transcatheter closure attempt: transesophageal echocardiography-planned device type and size were modified in 32 patients (64%). Rates of procedural success, predischarge occlusion, and major complications rate were 100%, 90%, and 2%, respectively. On mean follow-up of 5.4 ± 1.8 years, the follow-up occlusion rate was 98%. During follow-up, only one case of permanent atrial fibrillation was observed. There were no cases of aortic/atrial erosion, device thrombosis, or new atrioventricular valve dysfunction.
CONCLUSIONS: Intracardiac echocardiography-guided complex secundum ASD transcatheter closure is safe and effective and appears to have excellent long-term results, thus minimizing potential complications resulting from the complex anatomy.
© 2011 Wiley Periodicals, Inc.

Entities:  

Mesh:

Year:  2011        PMID: 22010972     DOI: 10.1111/j.1747-0803.2011.00567.x

Source DB:  PubMed          Journal:  Congenit Heart Dis        ISSN: 1747-079X            Impact factor:   2.007


  4 in total

1.  Utility of balloon assisted technique in trans catheter closure of very large (≥35 mm) atrial septal defects.

Authors:  Ajith Ananthakrishna Pillai; Vidhyakar Rangaswamy Balasubramanian; Raja Selvaraj; Maheshkumar Saktheeswaran; Santhosh Satheesh; Balachander Jayaraman
Journal:  Cardiovasc Diagn Ther       Date:  2014-02

Review 2.  Comprehensive understanding of atrial septal defects by imaging studies for successful transcatheter closure.

Authors:  Jinyoung Song
Journal:  Korean J Pediatr       Date:  2014-07-23

3.  Clinical evaluation of percutaneous and intra-operative device closure of atrial septal defects under transesophageal echocardiographic guidance: one center experience and mid-term follow-up.

Authors:  Peng Zhu; Haifeng Qiang; Fei Liu; Peng Xie; Shaoyi Zheng; Yong Sun
Journal:  J Cardiothorac Surg       Date:  2020-01-14       Impact factor: 1.637

4.  Trends in the Prevalence of Atrial Septal Defect and Its Associated Factors among Congenital Heart Disease Patients in Vietnam.

Authors:  Ho Xuan Tuan; Phan The Phuoc Long; Vu Duy Kien; Le Manh Cuong; Nguyen Van Son; Robert Dalla-Pozza
Journal:  J Cardiovasc Dev Dis       Date:  2019-12-27
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.