Literature DB >> 22009765

Tissue electrical properties monitoring for the prevention of pressure sore.

Congo Tak-Shing Ching1, Mei-Yun Chou, Siou-Jhen Jiang, Su-Hua Huang, Tai-Ping Sun, Wei-Hao Liu, Chia-Ming Liu.   

Abstract

BACKGROUND: Pressure sores are a significant problem in the healthcare sector. Although they may cause considerable morbidity, they are preventable.
OBJECTIVES: The objectives of this study are to (1) investigate the electrical properties of a tissue close to and away from the pressure sore site, and (2) establish a new approach for objective, reliable, low-cost and noninvasive screening or detection of pressure sore in its early stage. STUDY
DESIGN: Randomised controlled trial.
METHODS: Fifteen patients participated in this study. They all had stage I or stage II sacral pressure sores. Tiny surface electrodes in four-electrode configuration were used for all tissue electrical properties measurements recorded over the frequency range of 30-10 MHz.
RESULTS: Intraclass correlation coefficient (ICC) showed that all measurements (ICC > 0.90 for all measurements) had good reliability and validity. The real part of impedance (R) and the imaginary part of impedance (X) of a tissue measured close to the pressure sore site was found to be significantly smaller (p < 0.05 in all cases) than that measured away from the pressure sore site at a specific frequency range (R: 30.00-38.55 Hz; X: 43.95-606.40 Hz). It was also found that the extracellular resistance (R(e)) and the ratio of extracellular resistance to intracellular resistance (R(e)/R(i)) of a tissue measured close to the pressure sore site were significantly smaller (p < 0.05 in all cases) than that measured away from the pressure sore site.
CONCLUSIONS: Since the electrical properties (R, X, R(e), R(e)/R(i) ) of a tissue close to, and away from, the pressure sore site can be significantly distinguished, a potentially promising method for the screening of pressure sores at an early stage has been proposed.

Entities:  

Mesh:

Year:  2011        PMID: 22009765     DOI: 10.1177/0309364611424614

Source DB:  PubMed          Journal:  Prosthet Orthot Int        ISSN: 0309-3646            Impact factor:   1.895


  5 in total

1.  Use of prophylactic silicone adhesive dressings for maintaining skin integrity in intensive care unit patients: A randomised controlled trial.

Authors:  Yun J Lee; Jung Y Kim; Woo Y Shin
Journal:  Int Wound J       Date:  2019-03       Impact factor: 3.315

2.  Phantom testing of the sensitivity and precision of a sub-epidermal moisture scanner.

Authors:  Lea Peko Cohen; Amit Gefen
Journal:  Int Wound J       Date:  2019-04-16       Impact factor: 3.315

3.  Subepidermal moisture detection of pressure induced tissue damage on the trunk: The pressure ulcer detection study outcomes.

Authors:  Barbara M Bates-Jensen; Heather E McCreath; Anabel Patlan
Journal:  Wound Repair Regen       Date:  2017-05-31       Impact factor: 3.617

4.  A machine learning algorithm for early detection of heel deep tissue injuries based on a daily history of sub-epidermal moisture measurements.

Authors:  Maayan Lustig; Dafna Schwartz; Ruth Bryant; Amit Gefen
Journal:  Int Wound J       Date:  2022-01-12       Impact factor: 3.099

Review 5.  Subepidermal moisture (SEM) and bioimpedance: a literature review of a novel method for early detection of pressure-induced tissue damage (pressure ulcers).

Authors:  Zena Moore; Declan Patton; Shannon L Rhodes; Tom O'Connor
Journal:  Int Wound J       Date:  2016-04-29       Impact factor: 3.315

  5 in total