Thoracic artificial lung impedance studies using computational fluid dynamics and in vitro models.

Rebecca E Schewe1, Khalil M Khanafer, Ryan A Orizondo, Keith E Cook.   

Abstract

Current thoracic artificial lungs (TALs) possess blood flow impedances greater than the natural lungs, resulting in abnormal pulmonary hemodynamics when implanted. This study sought to reduce TAL impedance using computational fluid dynamics (CFD). CFD was performed on TAL models with inlet and outlet expansion and contraction angles, θ, of 15°, 45°, and 90°. Pulsatile blood flow was simulated for flow rates of 2-6 L/min, heart rates of 80 and 100 beats/min, and inlet pulsatilities of 3.75 and 2. Pressure and flow data were used to calculate the zeroth and first harmonic impedance moduli, Z(0) and Z(1), respectively. The 45° and 90° models were also tested in vitro under similar conditions. CFD results indicate Z(0) increases as stroke volume and θ increase. At 4 L/min, 100 beats/min, and a pulsatility of 3.75, Z(0) was 0.47, 0.61, and 0.79 mmHg/(L/min) for the 15°, 45°, and 90° devices, respectively. Velocity band and vector plots also indicate better flow patterns in the 45° device. At the same conditions, in vitro Z (0) were 0.69 ± 0.13 and 0.79 ± 0.10 mmHg/(L/min), respectively, for the 45° and 90° models. These Z(0) are 65% smaller than previous TAL designs. In vitro, Z(1) increased with flow rate but was small and unlikely to significantly affect hemodynamics. The optimal design for flow patterns and low impedance was the 45° model.

Entities:  

Mesh:

Year:  2011        PMID: 22009316      PMCID: PMC3518865          DOI: 10.1007/s10439-011-0435-x

Source DB:  PubMed          Journal:  Ann Biomed Eng        ISSN: 0090-6964            Impact factor:   3.934


  11 in total

1.  Predicting membrane oxygenator pressure drop using computational fluid dynamics.

Authors:  Kenneth L Gage; Mark J Gartner; Greg W Burgreen; William R Wagner
Journal:  Artif Organs       Date:  2002-07       Impact factor: 3.094

2.  In vitro fluid mechanical effects of thoracic artificial lung compliance.

Authors:  John W McGillicuddy; Sean D Chambers; Darren T Galligan; Ronald B Hirschl; Robert H Bartlett; Keith E Cook
Journal:  ASAIO J       Date:  2005 Nov-Dec       Impact factor: 2.872

3.  Effect of artificial lung compliance on in vivo pulmonary system hemodynamics.

Authors:  Hitoshi Sato; John W McGillicuddy; Grant W Griffith; Amy M Cosnowski; Sean D Chambers; Ronald B Hirschl; Robert H Bartlett; Keith E Cook
Journal:  ASAIO J       Date:  2006 May-Jun       Impact factor: 2.872

4.  Cardiac output during high afterload artificial lung attachment.

Authors:  Jeongho Kim; Hitoshi Sato; Grant W Griffith; Keith E Cook
Journal:  ASAIO J       Date:  2009 Jan-Feb       Impact factor: 2.872

5.  The relationship between pulmonary system impedance and right ventricular function in normal sheep.

Authors:  Alexander S Kuo; Hitoshi Sato; Junewai L Reoma; Keith E Cook
Journal:  Cardiovasc Eng       Date:  2009-12

6.  In-parallel artificial lung attachment at high flows in normal and pulmonary hypertension models.

Authors:  Begum Akay; Junewai L Reoma; Daniele Camboni; Joshua R Pohlmann; John M Albert; Ayushi Kawatra; Ayanna D Gouch; Robert H Bartlett; Keith E Cook
Journal:  Ann Thorac Surg       Date:  2010-07       Impact factor: 4.330

7.  Design and development of an artificial implantable lung using multiobjective genetic algorithm: evaluation of gas exchange performance.

Authors:  Ichiro Taga; Akio Funakubo; Yasuhiro Fukui
Journal:  ASAIO J       Date:  2005 Jan-Feb       Impact factor: 2.872

8.  Flow vectorial analysis in an artificial implantable lung.

Authors:  Akio Funakubo; Ichiro Taga; John W McGillicuddy; Yasuhiro Fukui; Ronald B Hirschl; Robert H Bartlett
Journal:  ASAIO J       Date:  2003 Jul-Aug       Impact factor: 2.872

Review 9.  Ventilator-induced lung injury and multiple system organ failure: a critical review of facts and hypotheses.

Authors:  Frans B Plötz; Arthur S Slutsky; Adrianus J van Vught; Cobi J Heijnen
Journal:  Intensive Care Med       Date:  2004-06-24       Impact factor: 17.440

10.  Pulmonary artery hemodynamics in primary pulmonary hypertension.

Authors:  W K Laskey; V A Ferrari; H I Palevsky; W G Kussmaul
Journal:  J Am Coll Cardiol       Date:  1993-02       Impact factor: 24.094

View more
  5 in total

1.  Early in vivo experience with the pediatric continuous-flow total artificial heart.

Authors:  Jamshid H Karimov; David J Horvath; Nicole Byram; Gengo Sunagawa; Barry D Kuban; Shengqiang Gao; Raymond Dessoffy; Kiyotaka Fukamachi
Journal:  J Heart Lung Transplant       Date:  2018-03-30       Impact factor: 10.247

2.  Use of a low-resistance compliant thoracic artificial lung in the pulmonary artery to pulmonary artery configuration.

Authors:  Christopher N Scipione; Rebecca E Schewe; Kelly L Koch; Andrew W Shaffer; Amit Iyengar; Keith E Cook
Journal:  J Thorac Cardiovasc Surg       Date:  2013-02-10       Impact factor: 5.209

3.  Design and in vitro assessment of an improved, low-resistance compliant thoracic artificial lung.

Authors:  Rebecca E Schewe; Khalil M Khanafer; Aarthi Arab; Jeffrey A Mitchell; David J Skoog; Keith E Cook
Journal:  ASAIO J       Date:  2012 Nov-Dec       Impact factor: 2.872

4.  Hemodynamic design requirements for in-series thoracic artificial lung attachment in a model of pulmonary hypertension.

Authors:  Begum Akay; Julie A Foucher; Daniele Camboni; Kelly L Koch; Ayushi Kawatra; Keith E Cook
Journal:  ASAIO J       Date:  2012 Jul-Aug       Impact factor: 2.872

5.  Low-Resistance, Concentric-Gated Pediatric Artificial Lung for End-Stage Lung Failure.

Authors:  Alex J Thompson; Skylar Buchan; Benjamin Carr; Clinton Poling; McKenzie Hayes; Uditha Piyumindri Fernando; Andreas Kaesler; Peter Schlanstein; Felix Hesselmann; Jutta Arens; Joseph A Potkay; Alvaro Rojas-PeÑa; Robert H Bartlett; Ronald B Hirschl
Journal:  ASAIO J       Date:  2020-04       Impact factor: 3.826

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.