Use of electron-beam sterilized hemodialysis membranes and risk of thrombocytopenia.

Mercedeh Kiaii1, Ognjenka Djurdjev, Myriam Farah, Adeera Levin, Beverly Jung, Jennifer MacRae.   

Abstract

CONTEXT: Thrombocytopenia is not widely recognized as a potential dialyzer-related complication. Following the observation of significant thrombocytopenia among 20 patients undergoing hemodialysis in a single dialysis unit after the introduction of dialyzers sterilized by electron beam (e-beam), a larger investigation was undertaken.
OBJECTIVE: To determine the prevalence and etiology of thrombocytopenia in hemodialysis populations of 2 Canadian provinces (British Columbia and Alberta). DESIGN, SETTING, AND PARTICIPANTS: A cohort study of patients undergoing hemodialysis in British Columbia (n = 1706) and southern Alberta (n = 425) between April 1, 2009, and November 30, 2010. Retrospective analyses of historical patient, laboratory, and dialyzer data predating conversion to e-beam dialyzers were undertaken, with prospective collection of predialysis and postdialysis platelet counts before and after the change from e-beam to non-e-beam sterilized dialyzers in September 2009. MAIN OUTCOME MEASURE: Significant thrombocytopenia, defined a priori as postdialysis treatment platelet count of less than 100 × 10(3)/μL and a postdialysis decrease in platelet count of more than 15%.
RESULTS: Among 1706 patients undergoing hemodialysis in British Columbia, 1411 (83%) were undergoing hemodialysis with e-beam sterilized dialyzers. Of 1706 patients, 194 (11.4%; 95% CI, 9.9%-12.9%) had postdialysis platelet counts of less than 100 × 10(3)/μL; 400 (23.4%; 95% CI, 21.5%-25.5%) had postdialysis decreases in platelet counts of more than 15%; and 123 (7.2%; 95% CI, 6.0%-8.6%) met both criteria. Among 425 patients in Alberta undergoing dialysis with polysulfone, e-beam sterilized dialyzers made by a different manufacturer, 46 (10.8%; 95% CI, 8.1%-14.3%) had platelet counts of less than 100 × 10(3)/μL; 156 (32.0%; 95% CI, 27.6%-36.7%) had decreases in platelet counts of more than 15%; and 31 (7.3%; 95% CI, 5.1%-10.3%) met both criteria. In multivariable analysis adjusting for patient and dialysis history characteristics, a significant association was observed between using an e-beam sterilized dialyzer and risk of significant thrombocytopenia (odds ratio [OR], 2.52; 95% CI, 1.20-5.29; P = .02). Compared with use of e-beam sterilized dialyzers, following the change to use of non-e-beam sterilized dialyzers, among 1784 patients, significant reductions were observed in postdialysis thrombocytopenia (120 patients [6.7%; 95% CI, 5.6%-8.0%; P < .001] had platelet counts of <100 × 10(3)/μL; 167 patients [9.4%; 95% CI, 8.1%-10.8%; P < .001] had decreases in platelet counts of >15%; and 38 patients [2.1%; 95% CI, 1.5%-2.9%; P < .001] met both criteria). Using generalized estimating equation modeling for repeated data with binary outcome, after adjusting for patient characteristics, the odds of significant thrombocytopenia were higher during the use of e-beam sterilized dialyzers than with use of non-e-beam sterilized dialyzers (OR, 3.57; 95% CI, 2.54-5.04; P < .001).
CONCLUSION: In this cohort of patients undergoing hemodialysis in 2 Canadian provinces in 2009-2010, the use of e-beam sterilized dialyzers was associated with significant thrombocytopenia following dialysis.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 22009100     DOI: 10.1001/jama.2011.1499

Source DB:  PubMed          Journal:  JAMA        ISSN: 0098-7484            Impact factor:   56.272


  14 in total

1.  Dialysis: Dialyzer sterilization method linked to thrombocytopenia.

Authors:  Susan J Allison
Journal:  Nat Rev Nephrol       Date:  2011-11-28       Impact factor: 28.314

2.  Dialysis-associated allergic reactions during continuous renal replacement therapy and hemodialysis: A case report.

Authors:  Dhruti P Chen; Jennifer E Flythe
Journal:  Hemodial Int       Date:  2019-11-19       Impact factor: 1.812

3.  Renal Association Clinical Practice Guideline on Haemodialysis.

Authors:  Damien Ashby; Natalie Borman; James Burton; Richard Corbett; Andrew Davenport; Ken Farrington; Katey Flowers; James Fotheringham; R N Andrea Fox; Gail Franklin; Claire Gardiner; R N Martin Gerrish; Sharlene Greenwood; Daljit Hothi; Abdul Khares; Pelagia Koufaki; Jeremy Levy; Elizabeth Lindley; Jamie Macdonald; Bruno Mafrici; Andrew Mooney; James Tattersall; Kay Tyerman; Enric Villar; Martin Wilkie
Journal:  BMC Nephrol       Date:  2019-10-17       Impact factor: 2.388

Review 4.  Maintaining safety in the dialysis facility.

Authors:  Alan S Kliger
Journal:  Clin J Am Soc Nephrol       Date:  2014-11-06       Impact factor: 8.237

5.  Thrombopoietic status of patients on haemodialysis.

Authors:  Taha Bat; Betul E Bat; Ahmed El-Moghraby; Samir Patel; Xingmin Feng; Cynthia E Dunbar; Erdal Sarac
Journal:  Br J Haematol       Date:  2016-02-16       Impact factor: 6.998

6.  Ability of a novel system for neonatal extracorporeal renal replacement therapy with an ultra-small volume circuit to remove solutes in vitro.

Authors:  Saeko Nishimi; Ken Ishikawa; Makoto Sasaki; Hiromi Furukawa; Akira Takada; Shoichi Chida
Journal:  Pediatr Nephrol       Date:  2015-10-22       Impact factor: 3.714

7.  Dialysis-related thrombocytopenia: a case report.

Authors:  Irene Faria Duayer; Maria Júlia Correia Lima Nepomuceno Araújo; Camila Hitomi Nihei; Maria Alice Fernandes Barcelos; Osni Braga; Zita Maria Leme Britto; Rosilene Mota Elias
Journal:  J Bras Nefrol       Date:  2022 Jan-Mar

8.  Recurrent white thrombi formation in hemodialysis tubing: a case report.

Authors:  Kiran P Sathe; Wee-Song Yeo; Isaac Desheng Liu; Sudha Ekambaram; Mohammed Azar; Hui-Kim Yap; Kar-Hui Ng
Journal:  BMC Nephrol       Date:  2015-01-15       Impact factor: 2.388

9.  Effectiveness of Haemodiafiltration with Heat Sterilized High-Flux Polyphenylene HF Dialyzer in Reducing Free Light Chains in Patients with Myeloma Cast Nephropathy.

Authors:  Mathieu Rousseau-Gagnon; Mohsen Agharazii; Sacha A De Serres; Simon Desmeules
Journal:  PLoS One       Date:  2015-10-14       Impact factor: 3.240

10.  Thrombocytopenia in the setting of hemodialysis using biocompatible membranes.

Authors:  Kathryn B Muir; Clifford D Packer
Journal:  Case Rep Med       Date:  2012-10-09
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.