Literature DB >> 22008329

Comparison of bipolar electrosurgical devices with ligatures and endoclips in the rat appendicitis model.

Levent Elemen1, Yusufhan Yazir, Alper Akay, Zehra Boyacıoğlu, Bahar Ceyran, Süreyya Ceylan.   

Abstract

PURPOSE: The aim of this study is to compare bipolar radiofrequency-driven vessel sealer, bipolar electrocautery, polyglactin 910 sutures, and endoclips in appendiceal stump closure with respect to operative time, appendiceal stump strength, and inflammation in a rat appendiceal model.
METHODS: Forty-eight Wistar-Albino rats, which had previously created appendicitis, were divided into 2 (groups A and B). Each group was further subdivided into 4 subgroups (AL, ABPC, AC, AS, BL, BBPC, BC, and BS) each containing 6 rats. An appendectomy was performed, and the stump was closed by bipolar radiofrequency-driven vessel sealer in the L, bipolar electrocautery in the BPC, endoclips in the C, and ligatures in the S subgroups. Cecum bursting pressures were determined instantly after the operation in group A and on the seventh postoperative day in group B. All operative times were measured. Appendices and appendiceal stumps underwent histopathologic examination. Statistical analyses were performed with Kruskal-Wallis and Mann-Whitney U tests. P < .05 was considered significant.
RESULTS: Bursting pressures of the subgroups were comparable on postoperative immediate period and day 7. Operative times were significantly shorter in the L and BPC subgroups in A and B. Histopathologic examination showed that the inflammation scores were similar in group A. In group B, inflammation parameters were also similar except the necrosis status, which was found to be decreased in BL compared with BC and BS. Necrosis status was significantly lower, and lymphocyte quantity was significantly higher in BL and BBPC compared with AL and ABPC.
CONCLUSIONS: Bipolar radiofrequency-driven vessel sealer and bipolar electrocautery achieve safe stump closure with satisfactory bursting pressure values in an experimental rat appendicitis model. Decreased operative time and unimpaired healing are encouraging.
Copyright © 2011 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2011        PMID: 22008329     DOI: 10.1016/j.jpedsurg.2011.04.062

Source DB:  PubMed          Journal:  J Pediatr Surg        ISSN: 0022-3468            Impact factor:   2.545


  4 in total

Review 1.  Non-vascular experimental and clinical applications of advanced bipolar radiofrequency thermofusion technology in the thorax and abdomen: a systematic review.

Authors:  Shobhit Arya; Hugh Mackenzie; George B Hanna
Journal:  Surg Endosc       Date:  2014-11-19       Impact factor: 4.584

2.  Comparison and efficacy of LigaSure and rubber band ligature in closing the inflamed cecal stump in a rat model of acute appendicitis.

Authors:  Chun-Chieh Yeh; Chia-Ing Jan; Horng-Ren Yang; Po-Han Huang; Long-Bin Jeng; Wen-Pang Su; Hui-Chen Chen
Journal:  Biomed Res Int       Date:  2015-01-28       Impact factor: 3.411

3.  Ex-Vivo Evaluation of "First Tip Closing" Radiofrequency Vessel Sealing Devices for Swine Small Intestinal Transection.

Authors:  Luca Lacitignola; Alberto Crovace; Giuseppe Passantino; Francesco Staffieri
Journal:  Vet Sci       Date:  2022-08-19

4.  Development of a standardized laparoscopic caecum resection model to simulate laparoscopic appendectomy in rats.

Authors:  Philipp Lingohr; Jonas Dohmen; Hanno Matthaei; Timo Schwandt; Gun-Soo Hong; Nils Konieczny; Edwin Bölke; Sven Wehner; Jörg C Kalff
Journal:  Eur J Med Res       Date:  2014-06-17       Impact factor: 2.175

  4 in total