Validation of a polyvinylidene fluoride impedance sensor for respiratory event classification during polysomnography.

Brian B Koo1, Colin Drummond, Susan Surovec, Nathan Johnson, Stephanie A Marvin, Susan Redline.   

Abstract

STUDY
OBJECTIVES: The AASM has recommended specific sensors in measuring apnea and hypopnea based on published reliability and validity data. As new technology emerges, these guidelines will need revision. Polyvinylidene fluoride (PVDF) measures impedance and can be incorporated into a belt to approximate airflow and respiratory effort. We compared respiratory event detection using PVDF impedance belts (PVDFb), respiratory inductance plethysmography (RIP), and nasal-oral pneumotachography (PNT).
METHODS: First, in a clinical setting, 50 subjects (median AHI 26) undergoing polysomnography were fitted with PVDFb and standard sensors. Studies were scored in 4 independent passes using 4 respiratory montages (M); M1: nasal pressure transduction (NPT), thermistry, and RIP; M2: NPT, thermistry, and PVDFb; M3: thermistry and PVDFb; M4: PVDFb alone. Each experimental montage (M2-M4) was compared to the reference standard (M1) for total apneas and hypopneas. In a second experimental study, respiratory event detection was compared across a series of breathing trials for PVDFb, RIP, and PNT in normal subjects. Agreement was evaluated with intraclass correlation coefficient (ICC), κ statistics, and Bland-Altman plots.
RESULTS: ICCs comparing event numbers by M1 to M 2, 3, and 4 were: 0.99, 0.93, and 0.91, respectively. Almost identical numbers of events were identified for M 1 and M2 (177.5 ± 122.7 vs 177.6 ± 123.2). Event subtypes also were comparable. PVDFb was less sensitive than PNT but no different than RIP in detecting decreased breathing amplitude.
CONCLUSIONS: PVDFb was comparable to standard RIP in determining respiratory events during polysomnography and in detecting decreased breathing amplitude, suggesting that PVDFb can be used as an alternative to RIP for apnea/hypopnea evaluation.

Entities:  

Keywords:  Sleep apnea; respiratory effort; signal; sleep; technology; validity

Mesh:

Substances:

Year:  2011        PMID: 22003343      PMCID: PMC3190847          DOI: 10.5664/JCSM.1312

Source DB:  PubMed          Journal:  J Clin Sleep Med        ISSN: 1550-9389            Impact factor:   4.062


  21 in total

1.  Comparison of direct and indirect measurements of respiratory airflow: implications for hypopneas.

Authors:  S Berg; J S Haight; V Yap; V Hoffstein; P Cole
Journal:  Sleep       Date:  1997-01       Impact factor: 5.849

2.  Relationship between inspiratory mouth pressure and respiratory muscle activity in normal subjects.

Authors:  G Y Ng; M J Stokes
Journal:  Respir Med       Date:  1992-07       Impact factor: 3.415

3.  IMPROVED TECHNIQUE FOR ESTIMATING PLEURAL PRESSURE FROM ESOPHAGEAL BALLOONS.

Authors:  J MILIC-EMILI; J MEAD; J M TURNER; E M GLAUSER
Journal:  J Appl Physiol       Date:  1964-03       Impact factor: 3.531

4.  Accuracy of thermistors and thermocouples as flow-measuring devices for detecting hypopnoeas.

Authors:  R Farré; J M Montserrat; M Rotger; E Ballester; D Navajas
Journal:  Eur Respir J       Date:  1998-01       Impact factor: 16.671

5.  Comparison of impedance and inductance ventilation sensors on adults during breathing, motion, and simulated airway obstruction.

Authors:  K P Cohen; W M Ladd; D M Beams; W S Sheers; R G Radwin; W J Tompkins; J G Webster
Journal:  IEEE Trans Biomed Eng       Date:  1997-07       Impact factor: 4.538

6.  Sleep-disordered breathing and cardiovascular disease: cross-sectional results of the Sleep Heart Health Study.

Authors:  E Shahar; C W Whitney; S Redline; E T Lee; A B Newman; F J Nieto; G T O'Connor; L L Boland; J E Schwartz; J M Samet
Journal:  Am J Respir Crit Care Med       Date:  2001-01       Impact factor: 21.405

7.  Transducer system for the noninvasive recording of arterial pressure contours.

Authors:  S G Karr; T Karwoski; J E Jacobs; L F Mockros
Journal:  Ann Biomed Eng       Date:  1985       Impact factor: 3.934

8.  The occurrence of sleep-disordered breathing among middle-aged adults.

Authors:  T Young; M Palta; J Dempsey; J Skatrud; S Weber; S Badr
Journal:  N Engl J Med       Date:  1993-04-29       Impact factor: 91.245

9.  History and recent progress in piezoelectric polymers.

Authors:  E Fukada
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2000       Impact factor: 2.725

10.  Obstructive sleep apnea syndrome, sleepiness, and quality of life.

Authors:  Marta A Goncalves; Teresa Paiva; Elizabeth Ramos; Christian Guilleminault
Journal:  Chest       Date:  2004-06       Impact factor: 9.410

View more
  5 in total

1.  A Potential Alternative to Respiratory Inductance Plethysmography for Children?

Authors:  Jason Z Bronstein; Lee J Brooks
Journal:  J Clin Sleep Med       Date:  2017-02-15       Impact factor: 4.062

2.  Validation of Polyvinylidene Fluoride Impedance Sensor for Respiratory Event Classification during Polysomnography in Children.

Authors:  Anne G Griffiths; Pallavi P Patwari; Darius A Loghmanee; Matthew J Balog; Irina Trosman; Stephen H Sheldon
Journal:  J Clin Sleep Med       Date:  2017-02-15       Impact factor: 4.062

3.  Rules for scoring respiratory events in sleep: update of the 2007 AASM Manual for the Scoring of Sleep and Associated Events. Deliberations of the Sleep Apnea Definitions Task Force of the American Academy of Sleep Medicine.

Authors:  Richard B Berry; Rohit Budhiraja; Daniel J Gottlieb; David Gozal; Conrad Iber; Vishesh K Kapur; Carole L Marcus; Reena Mehra; Sairam Parthasarathy; Stuart F Quan; Susan Redline; Kingman P Strohl; Sally L Davidson Ward; Michelle M Tangredi
Journal:  J Clin Sleep Med       Date:  2012-10-15       Impact factor: 4.062

4.  The use of combined thermal/pressure polyvinylidene fluoride film airflow sensor in polysomnography.

Authors:  Meir Kryger; Todd Eiken; Li Qin
Journal:  Sleep Breath       Date:  2013-05-29       Impact factor: 2.816

5.  Deep Recurrent Neural Networks for Automatic Detection of Sleep Apnea from Single Channel Respiration Signals.

Authors:  Hisham ElMoaqet; Mohammad Eid; Martin Glos; Mutaz Ryalat; Thomas Penzel
Journal:  Sensors (Basel)       Date:  2020-09-04       Impact factor: 3.576

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.