Literature DB >> 22001719

Results from an Australian and New Zealand audit of left ventricular ejection fraction from gated heart pool scan analysis.

Elizabeth A Bailey1, Dale L Bailey.   

Abstract

AIM: A voluntary audit was undertaken to compare left ventricular ejection fraction (LVEF) calculations from gated heart pool studies using software packages currently available throughout Australia and New Zealand (ANZ).
METHODS: A data set previously developed by a task group of the Institute of Physics and Engineering in Medicine (IPEM) consisting of 12 planar gated studies with LVEF ranging from 17 to 67% was provided to each site with worksheets to record individual technologist's results as well as information about individual department's equipment, software used for calculation of LVEF and experience of the analysers. Data analysis included calculation of an overall mean LVEF and standard deviation (SD) for all patients, which were compared with the mean IPEM LVEF. A comparison among software packages was done to examine differences in mean LVEF.
RESULTS: Results were received from a total of 22 sites using six different software packages, with 187 processed studies included for analysis. Reproducibility of studies 4 and 8 showed good agreement, with an SD of 1.0 EF units (P=0.5290). Good correlation was found between the ANZ mean LVEF and the corresponding IPEM value per patient study, with R=0.98. The ANZ overall mean LVEF was 47%, which was higher that the IPEM mean by 4.9 EF units. The average difference among all software packages was +5.1 EF units.
CONCLUSION: Differences were found between the IPEM mean and the ANZ mean LVEF. Variability in LVEF values between software packages was demonstrated, with an average of 5.1 EF units. The ANZ overall mean for LVEF was 47.2%, being 4.9 EF units higher than the IPEM value.

Entities:  

Mesh:

Year:  2012        PMID: 22001719     DOI: 10.1097/MNM.0b013e32834c2f0b

Source DB:  PubMed          Journal:  Nucl Med Commun        ISSN: 0143-3636            Impact factor:   1.690


  4 in total

1.  Comparison of estimates of left ventricular ejection fraction obtained from gated blood pool imaging, different software packages and cameras.

Authors:  Rachelle Steyn; John Boniaszczuk; Theodore Geldenhuys
Journal:  Cardiovasc J Afr       Date:  2014 Mar-Apr       Impact factor: 1.167

2.  Accuracy of left ventricular ejection fraction by contemporary multiple gated acquisition scanning in patients with cancer: comparison with cardiovascular magnetic resonance.

Authors:  Hans Huang; Prabhjot S Nijjar; Jeffrey R Misialek; Anne Blaes; Nicholas P Derrico; Felipe Kazmirczak; Igor Klem; Afshin Farzaneh-Far; Chetan Shenoy
Journal:  J Cardiovasc Magn Reson       Date:  2017-03-24       Impact factor: 5.364

Review 3.  Cardiac monitoring in HER2-positive patients on trastuzumab treatment: A review and implications for clinical practice.

Authors:  Nathalie I Bouwer; Agnes Jager; Crista Liesting; Marcel J M Kofflard; Jasper J Brugts; Jos J E M Kitzen; Eric Boersma; Mark-David Levin
Journal:  Breast       Date:  2020-04-16       Impact factor: 4.380

4.  NT-proBNP correlates with LVEF decline in HER2-positive breast cancer patients treated with trastuzumab.

Authors:  Nathalie I Bouwer; Crista Liesting; Marcel J M Kofflard; Sylvia M Sprangers-van Campen; Jasper J Brugts; Jos J E M Kitzen; Michael A Fouraux; Mark-David Levin; Eric Boersma
Journal:  Cardiooncology       Date:  2019-05-28
  4 in total